Function check
429
430
Function check
431
25
If you think your NovoPen 3 PenMate is not working properly, follow this procedure:
432
a) Check that the barrel and the NovoPen 3 PenMate are screwed together tightly and
433
that the dose indicator window on the barrel is aligned with the yellow push button on
434
the NovoPen 3 PenMate.
435
b)
Screw on a new NovoFine needle as described in step10.
436
c)
Clear the air bubbles as described in step11.
437
d) Put the outer needle cap over the needle.
438
e)
Dispense 20 units into the needle cap.
439
440
The insulin should fill the lower part of the outer needle cap. If the pen has delivered
441
too much or too little insulin, repeat the test. If it happens again, do not use the pen.
442
Contact your supplier.
443
Do not try to repair a faulty NovoPen 3 or a faulty NovoPen 3 PenMate.
444
Never use NovoPen 3 or NovoPen 3 PenMate unless you are sure that they are
445
working properly.
446
447
Summary of Contents for NovoPen 3 PenMate
Page 1: ...1 ...
Page 4: ...NovoPen 3 PenMate Instruction Manual Page 4 FDA revision 4 final 51 ...
Page 5: ...NovoPen 3 PenMate Instruction Manual Page 5 FDA revision 4 final 52 PenFill Holder ...
Page 18: ...NovoPen 3 PenMate Instruction Manual Page 18 FDA revision 4 final 272 273 274 275 ...
Page 20: ...NovoPen 3 PenMate Instruction Manual Page 20 FDA revision 4 final 289 ...
Page 22: ...NovoPen 3 PenMate Instruction Manual Page 22 FDA revision 4 final 302 ...
Page 29: ...NovoPen 3 PenMate Instruction Manual Page 29 FDA revision 4 final 390 391 392 ...
Page 34: ...NovoPen 3 PenMate Instruction Manual Page 34 FDA revision 4 final 448 ...