background image

LogBox-AA Data Logger 

NOVUS AUTOMATION 

2/6

 

NXPERIENCE SOFTWARE 

You must use 

NXperience

 software to configure and download data 

from LogBox-AA. To install the software, simply run the file 

NXperience_Setup.exe

, available on our website. 

RUNNING NXPERIENCE 

When you start 

NXperience

, the home screen will appear: 

 

Figure 3 –

 

NXperience 

home screen 

Next, at the bottom of the screen, you must click on the   button to 

indicate the serial port to be used by the 

IR-LINK3 Optical Interface

which must be connected to the computer USB: 

 

Figure 4 –

 

Select a device 

In the example above you should select the option "USB Serial Port 

(COM3)" and click 

Ok

. After that, simply click on the 

Configure

 

button and then on 

Read Device

 to perform the connection process. 

Note:

 This manual provides basic information on how to configure 

the device. For more specific information about  other features, 

please refer to the 

NXperience

 manual, available on our website. 

IR-LINK3 OPTICAL INTERFACE 

To configure, monitor, or download data from the logger through the 

NXperience

 software, you must use an 

IR-LINK3 Communication 

Interface

 (purchased separately).  

IR-LINK3

  acts as an intermediary between the device and the 

computer being used and allows you to transmit and receive data 

from the logger via infrared signals. 
The driver will be installed during the 

NXperience

  installation 

process. Once this is done, 

IR-LINK3

 will be recognized whenever it 

is connected to the computer. 

CONFIGURING THE LOGGER 

To configure the logger, 

IR-LINK3 Communication Interface

 must 

be connected to the computer being used.  

IR-LINK3

  must be directed constantly towards the logger 

communication window (see 

Figure 5

) and kept at approximately 15 

cm: 

 

Figure 5 –

 

Interface directed to the logger 

According to the instructions defined in the 

NXPERIENCE 

SOFTWARE

 item, you must click the 

Read Device

 button to start the 

communication between the recorder and the software. Then, the 

general settings screen will be presented: 

 

Figure 6 –

 

General settings 

This screen displays general information about the device, such as 

serial number, firmware version, and model. In the 

Device Tag

 

parameter, you can set a specific name for 

LogBox-AA

If the communication between the logger and the computer is 

established,  the date and time displayed on the screen will be 

constantly updated. 

Afterwards, if you need to return to this screen, just click on the 

button. 

CHANNELS 

By clicking the 

 button, you can access the configuration fields of 

the logger analog channels: 

 

Figure 7 –

 

Analog channel 

The configuration fields for the analog channels are the following: 

Tag

: Set of up to 8 characters that allows you to identify the variable. 

Input Type

:  Allows you to define the signal applied to the analog 

input of the logger. The list shows all the available input options. The 

option adopted must be according to the jumper configuration (see 

INPUT CONNECTIONS

 item). 

Unit

:  Allows you to define the unit to be used for the measured 

variable. 

Mode:

 Allows you to define the recording mode:

 

Instant:

  The device performs a measurement. The value is 

recorded at each acquisition "Interval".

 

Medium:

 The device periodically takes ten measurements over 

each acquisition "Interval". The recorded value is the average of 

these measurements.

 

Minimum:

 The device periodically takes ten measurements over 

each acquisition "Interval". The recorded value corresponds to the 

smallest measurement.

 

Maximum:

 The device periodically takes ten measurements over 

each acquisition "Interval". The recorded value corresponds to the 

largest measure. 

Summary of Contents for LogBox-AA

Page 1: ...N R S or B voltage 0 to 50 mV or 0 to 10 V or current 0 to 20 mA or 4 to 20 mA depending on the configuration made Note In addition to performing the configuration via software you must properly position the internal jumper DATA ACQUISITION LOGGING The logger can 1 Perform a single measurement within a defined period and log the value read or 2 perform ten measurements in this period and log in me...

Page 2: ... and kept at approximately 15 cm Figure 5 Interface directed to the logger According to the instructions defined in the NXPERIENCE SOFTWARE item you must click the Read Device button to start the communication between the recorder and the software Then the general settings screen will be presented Figure 6 General settings This screen displays general information about the device such as serial nu...

Page 3: ... START MODE Immediate Start Allows to start the acquisition process right after sending the configuration to the logger Date Time Allows to start the acquisitions at a defined date and time Digital Input Allows to perform acquisitions while the digital input is triggered closed and to stop acquisitions when the digital input is disabled open Start Button Allows to start the acquisitions using the ...

Page 4: ...ve the same input connection scheme CN1 Pt100 6 1 2 3 4 5 Pt100 CN1 6 1 2 3 4 5 4 20mA 4 20mA Before using the logger you must set the internal jumpers according to the input type used The factory setting of these jumpers is to measure Pt100 Thermocouple 0 50 mV signals The figures below show the positions for the possible input types used INPUT SIGNAL CHANNEL 1 J1 POSITION CHANNEL 2 J2 POSITION 4...

Page 5: ... of the graph there are two tabs Data View and General Information Through the toolbar at the bottom of the download screen NXperience allows you to save save to file merge graphs filter and even export this information to several types of reports Figure 16 Toolbar Specific information about each of these features can be found in the NXperience manual available on our website DATA VIEW Figure 17 D...

Page 6: ... RTD s NBR 13773 97 TYPE FEATURES T C J Range 50 to 760 C 58 to 1400 F T C K Range 90 to 1370 C 130 to 2498 F T C T Range 100 to 400 C 148 to 752 F T C N Range 90 to 1300 C 130 to 2372 F T C R Range 0 to 1760 C 32 to 3200 F T C S Range 0 to 1760 C 32 to 3200 F T C B Range 150 to 1820 C 32 to 3308 F Pt100 Range 200 0 to 650 0 C 328 to 1202 F 0 50 mV Linear Programmable range of 32768 to 32767 4 20 ...

Reviews: