background image

TagTemp-NFC-LCD 

 

NOVUS AUTOMATION 

11/13

 

The "

Channels

" screen displays the following parameters: 

Input Type

: Displays the type of sensor in the device. 

Unit

: Sets the unit for the monitored quantity: 

°C

  or 

°F

  for the 

temperature channel.

 

Lower and Upper Limit

: Displays sensor limits.

 

Lower  and Upper  Alarm

: Allows to define the limits of values, 

which, when exceeded, characterize an alarm condition. The alarm 

events are indicated to the user through the 

Alarm Indicator

.

 

Offset

: Allows to make adjustments to the registered values. 

 

Fig. 32

 – Digitals Channels Screen 

Note

:  This screen is only displayed when reading 

TagTemp-NFC-

LCD 

model which has a digital input. In order to enable the digital 

input, please slide the red button. 

 

Fig. 33 

– Data Logs Screen 

The "

Logs

" screen displays the following parameters: 

Interval between logs

: sets the time interval between data 

acquisitions. The minimum interval is 1 minute and the maximum is 

up to 240 minutes. 

Estimated time

: Informs how long it will take to fill the available 

memory, with the conditions defined in the configuration. 

Start logs

: The recording of data can begin in two ways: “Immediate 

Start” or by using the device “By button”. 

Stop logs

: The logging can be stopped in two ways, when the 

memory is completed or by pressing the button on the device. Both 

modes can be configured to operate independently. If the option is 

set to "Do not Stop", the registration is done in a circular way, in 

which, when the memory is full, the oldest records will be overwritten 

with the new values. 

Maximum and Minimum: 

This option resets the maximum and 

minimum values when a configuration is sent. 

 

 

Fig. 34

 – Finalization Screen

 

The "

Send Configuration and Extras

" screen is where the 

configuration settings are sent and allows you to perform some other 

useful functions. 

Send Configuration: 

This button has three modes, Send 

Configuration, Send Configurations and Start Logs, or Stop Logging. 

In all three modes the device must be present in the NFC reader. 

Save file

: This option allows you to save the configuration settings 

made to a file, so it can be reused to configure other devices. 

Security

: This option allows you to create, change or delete a 

password. The password must be four digits long. 

Fig. 34

 shows the 

definition and password change screen. 

Reset the Minimum and Maximum: 

This button resets the device's 

minimum and maximum values.

 TagTemp-NFC-LCD 

must be in the 

NFC reader field for the action to be performed. 

  

Download logs

: This option allows the data in the device memory to 

be transferred to the application. For more information, see 

"

Downloading Data

".

 

Stop Loggings

: This option is only displayed when the configuration 

is being read and the device is logging, so that to send a new 

configuration, you need to stop the current log. See the following 

image. 

 

Fig. 35 

– Finalization - Stop loggings 

10.4

 

Downloading and Viewing Data 

The data download transfers the values recorded by the device to the 

computer. The download can occur at any time, during or after the 

completion of the data acquisition process. If data download occurs 

during the acquisition process, this process 

is not interrupted

, and 

the operation is set according to the device configuration settings. 

Summary of Contents for TagTemp-NFC-LCD

Page 1: ...ON 1 13 TagTemp NFC LCD OPERATION MANUAL V1 0x CE Mark This is a Class A product In a domestic environment this product may cause radio interference in which case the user may be required to take adeq...

Page 2: ...want to measure the temperature Fig 01 Screw fastening Fig 02 Frame 3 TECHNICAL DATA Measure Range Temperature 40 0 C to 100 0 C Operating temperature From 10 C to 60 C Accuracy 1 C max over the full...

Page 3: ...display the minimum and maximum temperature values reset the minimum and maximum values when this option is configured and start and or stop the records depending on the configuration chosen To start...

Page 4: ...ent has occurred 6 3 Log Signaling LOG Remains lit while the device is logging or blinking every 1s when waiting for the Start Delay time 6 4 Low Battery Signaling Indicates that the battery is low vo...

Page 5: ...s Logging State Displays whether the device is recording data is stopped or is waiting for the start of the records Start Delay or overwriting old records Memory Usage Indicates if the circular buffer...

Page 6: ...record this occurrence to the memory Use Fahrenheit degrees when this option is enabled the temperature unit of Celsius default is changed to Fahrenheit Reset Minimum and Maximum Resets the minimum an...

Page 7: ...r limit of range Average Temperature within Occurrences Average value of the temperatures above the maximum range Low Excursions Indicates when the measured value exceeds the lower limit under the spe...

Page 8: ...aph display range the Chart range option can be used 9 2 5 1 Data Chart Options Screen When you select the icon in the upper right of the screen the chart options will appear Fig 19 Data Chart Options...

Page 9: ...application Initial screen selection Allows you to select the startup screen after reading the data from the device NOVUS Cloud Authentication Hire an account in the NOVUS Cloud to store the TagTemp...

Page 10: ...out the need of the device This setting can be saved to a file for future use or recorded to a connected device The Open Configuration option is used to read an existing configuration file The reading...

Page 11: ...with the new values Maximum and Minimum This option resets the maximum and minimum values when a configuration is sent Fig 34 Finalization Screen The Send Configuration and Extras screen is where the...

Page 12: ...1 Sharing and processing data After completing the download a screen will be displayed with the downloaded data and these are presented in two forms in graph form and table form In the same screen th...

Page 13: ...roperties 9 Reports Export Logs 10 Sending data to NOVUS Cloud 11 Exit 10 5Exporting and creating reports with the data downloaded Using the NXperience you can export the downloaded data to various fi...

Reviews: