NOVUS AUTOMATION 

4/6

 

4 -  Channels

: Displays other parameters related to temperature 

measurement: 

 

Fig. 06

 – Configuration window

 

Unit

: Defines the unit of the value measured: 

ºC

  or 

ºF

  for 

channel 1 (temperature). 

Value

: It defines how the value measured will be registered. 

Options are:  

 

Instantaneous

: The value logged will be exactly the value 

measured at each interval defined. Measurement is taken 

at the end of the defined interval. The minimum interval 

between measurements is 5 seconds. 

 

Average

: The registered value will be the average of 10 

measurements made on logging interval.  The minimum 

time between each record is 50 seconds. 

 

Maximum

: The value logged will be the maximum value 

found within 10 consecutive measurements taken within a 

predefined interval. The minimum interval between 

measurements is 50 seconds. 

 

Minimum

: The value logged will be the minimum value 

found within 10 consecutive measurements taken within a 

predefined interval. The minimum interval between 

measurements is 50 seconds. 

Offset

:

 

Makes possible to correct the value logged. 

Alarm

: Defines limit values that, when exceeded, characterize 

an  alarm condition. Alarm events are informed to the user 

through a flashing mode of the 

Alarm Indicator

After filling all the fields, select 

OK

 and the settings will be sent to the 

device. 

4.4

 

STOP / PAUSE / RESUME BUTTONS 

These buttons will be enabled once the device is fully configured. 

The

 Stop 

button allows stop records permanently, so that the device 

will only register again after a new configuration by the user. 

 

The 

Pause

 / 

Resume

 button action sends the appropriate command, 

depending on the current registration status: 
If the records are 

In Progress

  and pressing this button, it will be 

send a command to 

Pause

 the records. 

 

Pause

: Pauses the registry, allowing it to be resumed later using the 

Resume

 button. 

If registration is 

Paused

 and pressing this button, it will be send a 

command to 

Resume

 the records. 

 

Resume

: Resumes the records that were interrupted by the 

Pause

 

command without erasing the previous registers of the device 

memory, using the same parameter settings previously configured. 

5

 

DOWNLOAD AND DATA VISUALIZATION 

Downloaded  data  are send from  device  to PC. Data can be 

downloaded any time at the end of the acquisition process or while 

they are being acquired. If data downloaded takes place during the 

acquisition process, the process will not be interrupted, following the 

device configuration. 

5.1

 

DOWNLOADING DATA 

Data  download  is accomplished by clicking the button 

Download 

Data

 

During data transference, a status bar indicates remaining data to be 

transferred. Data downloading time is proportional to the number of 

readings logged. 

5.2

 

VISUALIZING DATA 

At the end of values transfer, data can be displayed in a graphical 

form. 

5.2.1

 

Chart Window 

It is possible to select a region of the chart to zoom in. Zoom 

commands can be accessed through the 

View

  menu or through 

zoom icons from the task bar. 
It is also possible to select an area from the chart to zoom in by 

clicking and dragging the mouse, thus creating a zoom region 

starting from the upper left corner on the chart area. 
The graphic curves can be vertically dragged with a right-click and 

dragging the mouse up and down. 

Summary of Contents for TagTemp-USB

Page 1: ...agTemp USB TEMPERATURE LOGGER INSTRUCTIONS MANUAL V1 0x E CE Mark This is a Class A product In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures ...

Page 2: ...ion and is performing logs in memory 2 TECHNICAL DATA Measure Range Temperature 20 0 C to 70 0 C Accuracy 0 5 C 25 ºC 1 C max along the entire measured range Note A measure error can be fixed using the OFFSET parameter in the LogChart II software Measurement Resolutions Temperature 0 1 C Memory capacity 32 000 32 k logs Measurement Interval 5 seconds min 18 hours max Supply 3 0 V lithium battery C...

Page 3: ...e device Fig 05 Configuration window Fields are 1 Title In this field the user identifies the device by assigning it a name 2 General information Area with information about the device such as Model Serial Number Data Logger Date Hour Actual Date Hour Firmware Version Memory Capacity and Number of acquisitions stored in memory In this field time is constantly updated while the communication betwee...

Page 4: ...ation by the user The Pause Resume button action sends the appropriate command depending on the current registration status If the records are In Progress and pressing this button it will be send a command to Pause the records Pause Pauses the registry allowing it to be resumed later using the Resume button If registration is Paused and pressing this button it will be send a command to Resume the ...

Page 5: ...n can be displayed by pressing the icon Parameter Viewing Fig 09 General information window 5 3 EXPORTING DATA Downloaded data can be exported to different file formats for further analysis To export access File Export or click the icon 5 4 DIAGNOSTICS Battery Level Indicates the level of battery voltage When the battery reaches a low level voltage immediate replacement is recommended to ensure th...

Page 6: ...e device fails Make sure the COM port is selected correctly and there is no other program using the same port during communication attempts Make sure there is no physical obstacle blocking the infrared signal Make sure the cable is well connected to the PC port Make sure the port selected does not present any problem 8 SPECIAL CARE When handling the device and its battery observe the following Whe...

Reviews: