Interval

:  Defines  the  interval  between  readings:  hour,  minutes 

and  seconds.  When  the  logging  mode  is  set  to  the 

instantaneous

 the minimum interval is 1 second. For 

maximum

 

and 

minimum

  values,  the  minimum  interval  is  10  seconds  in 

both models. 

Misc

:  Shows  the  Logging  Mode  parameter  that  defines  how 

measured value will be logged. Options are:

 

Instantaneous

:  The  value  logged  will  be  exactly  the  value 

measured at each interval. Measurement is taken at the end 
of  the  defined  interval.  The  minimum  interval  between 
measurements is 1 second. 

Minimum

:  The  value  logged  will  be  the  minimum  value 

found within ten consecutive measurements taken within an 
interval. The minimum interval between measurements is 10 
second. 

Maximum

:  The  value  logged  will  be  the  maximum  value 

found within ten consecutive measurements taken within an 
interval. The minimum interval between measurements is 10 
second.

 

Alarms

:  Enables  an  alarm  that  is  triggered  according  to  user-

defined  parameters.  Values  shown  in  the  temperature  unit 
selected. 

 

Figure 14 – PDA Alarm Settings screen 

Clocks

:  Provides  access  to  Logger  and  PDA  clocks.  When  a 

new configuration is sent to the logger, clocks are updated. 

Offset

: allows fine offset adjustment on the measured value. The 

Reset

  button  clears  changes  made  and  the  logged  values  are 

the measured values again. 

 

Figure 15 – Offset adjust screen 

After configuring clocks in the 

Settings

 screen, click 

Apply

 to send 

this configuration to the Logger, returning to the 

Monitoring

 screen. 

Sending  a  new  configuration  implies  the 

exclusion

  of  all  data 

collected present in the logger memory.

 

7.3

 

DOWNLOADING DATA FROM THE LOGGER 

In  the 

Monitoring

  screen,  the 

Download

  button  performs  the 

transfer  of  the  data  from  de  LogBox  to  the  PDA.  Download  can  be 
partial and it does not interfere in the ongoing acquisition process. 

The data base of loggings is displayed in the 

Recorded Data

 screen, 

identified with the name assigned to the process (

Data Title

) and the 

date it was downloaded. 

Should the PDA batteries be discharged, all readings will be lost.

 

7.4

 

FILE VISUALIZATION 

The 

Recorded Data

 screen lists the data base logged and stored in 

the  PDA.  To  access  data,  select  the  desired  data  base  and  press 

Details

Recorded  Data  Details

  screen  shows  several  information 

about the data base. 

View Data

 shows in table format the logged values and the date and 

time they were performed. 

Press 

Delete

 to erase the selected data base. 

7.5

 

TRANSFERRING DATA TO YOUR DESKTOP 

HotSync

  of  data  stored  in  a  PDA  with  the  Desktop  is  performed 

through a conduit installed together with the LogChart Palm-OS. The 
conduit converts the data collected by the LogChart Palm-OS to a file 
compatible with the LogChart-II software. 

To access the conduit options, the 

HotSync Manager

 software must 

be  active.  Click  on  the  HotSync  Manager  in  the  Windows  taskbar. 
Select  in  the  drop-down  menu  the  option 

Custom.

  Select 

Novus 

LogChart  Conduit

  and  click 

Change.

  The  following  window  will  be 

exhibited:

 

 

Figure 16 – LogChart Conduit options 

Handheld  overwrites  Desktop

LogChart  Palm-OS

  files  are 

transferred to the Desktop. 

Save  in

:  Choose  a  directory  to  record  files  generated  during  data 

synchronization. 

Leave Data on Palm

: Option to keep or delete the data in the PDA 

after HotSync. 

Do nothing

: Data synchronization will not be performed; 

Set as default

: The same settings will be used in the next 

HotSync

 

processes. 

8

 

MOST FREQUENTLY FOUND PROBLEMS 

The  LED  is  not  flashing

:  The  LED  flashing  light  is  intentionally 

weak, and it can be difficult to see it in clear environments, certify it is 
really not flashing. 

Communication with the logger fails

:

 

Make sure the COM port is selected correctly and there is no other 
program using the same port during communication attempts; 

Make sure there is no physical obstacle blocking the infrared signal; 

Make sure the cable is well connected to the PC port; 

Make sure the port selected does not present any problem; 

9

 

SPECIAL CARE 

When handling the device and its battery, observe the following: 

 

When  the  device  is  opened,  avoid  touching  the  electronic 
components  and  the  printed  circuit  board  to  avoid  electrostatic 
discharge damages. 

 

Pay close attention to battery polarity. 

 

When  closing,  the  cover  and  its  sealing  must  be  mounted 
correctly and carefully to ensure the enclosure sealing. 

Summary of Contents for TagTemp

Page 1: ...Warns the user as to alarm conditions Whenever an alarm situation takes place it will flash once at every four seconds until a new configuration is applied to the logger 2 TECHNICAL DATA Measure Range Temperature 20 0 C to 70 0 C Accuracy 0 5 ºC 25 ºC 2 ºC max along the entire measured range Note A measure error can be fixed using the OFFSET parameter in the LogChart II software Measurement Resolu...

Page 2: ...neral information Area with information about the logger such as Model Serial Number Logger Date Time PC Date Time firmware version memory capacity and number of acquisitions stored in memory etc In this field time is constantly updated while the communication between logger and computer is taking place 3 Readings Presents a series of parameters that define the acquisition process Interval Defines...

Page 3: ...lable USB port Windows will request the proper driver installation which is found in the CD ROM that is provided with the logger 5 1 USB DRIVES INSTALLATION The drives installation steps may vary according to the machine even for the same version of an operating system The following screenshots and steps are only to provide guidance 1 Insert the CD shipped with the logger in the CD ROM drive 2 Con...

Page 4: ...ART WINDOW It is possible to select a region of the chart to zoom in Zoom commands can be accessed through the View menu or through zoom icons from the task bar It is also possible to select an area from the chart to zoom in by clicking and dragging the mouse thus creating a zoom region starting from the upper left corner on the chart area The graphic curves can be vertically dragged with a right ...

Page 5: ... a new Monitoring screen Download Offload logged data Download can be partial and it does not interfere in the ongoing measurement process More Info Displays further information on the connected logger such as model serial number version and memory capacity Settings Accesses the Settings screen which allows for modifying the logger configuration Data Base Displays the Recorded Data screen listing ...

Page 6: ...ownloaded Should the PDA batteries be discharged all readings will be lost 7 4 FILE VISUALIZATION The Recorded Data screen lists the data base logged and stored in the PDA To access data select the desired data base and press Details Recorded Data Details screen shows several information about the data base View Data shows in table format the logged values and the date and time they were performed...

Page 7: ...letrônicos Ltda products are covered by a 12 month warranty provided the purchaser presents the sales receipt and the following conditions are met Products are covered for one year from the original date of purchase Please retain the dated sales receipt as evidence of the date of purchase You will need it for any warranty service Within this period warranty against defects in material and workmans...

Reviews: