ISO-STIM 01D User Manual 

_________________________________________________________________________________________________________________

 

___________________________________________________________________________ 
version 4.3 

 

 

page 9 

(

22

)  SIGNAL INPUT connector 

 
BNC connector for the input of an analog signal in ANALOG or DIRECT mode (see 
also Chapter 4.2) 
 

 
(

23

)  ON / OFF switch 

 

Switch to turn POWER of the ISO-STIM 01D ON or OFF, respectively 
 

 

4.  Operation 

4.1. Connections – Noise Reduction 

 
The  ISO-STIM  01D  comes  with  a  power  supply  (13.8 V,  DC)  with  two  outputs  which  are 
connected to the stimulus isolation device by two cables with banana plugs. The corresponding 
jacks at the ISO-STIM 01D are located at the rear panel, or use the FILTERED/UNFILTERED 
switch. 
 

Important

: In most cases the noise level is substantially reduced, if the black cable from the 

power supply is connected additionally to ground!! 
 

Note

: Another possible source of noise is the chassis of the stimulator. In a noisy environment 

it  may  act  as  an  antenna  picking  up  50 Hz  or  60 Hz  noise  respectively,  especially  when 
operating with current stimuli. In this case the noise is substantially reduced if the chassis is 
connected to ground (see also #

11

, Figure 1). 

 

Note

: Use only the power supply shipped with the ISO-STIM 01D! 

 
The stimulation electrode is connected to isolated output (see Figure 1). We recommend to use 
a bipolar electrode for stimulating nerve bundles. This will produce only a short stimulation 
artifact at the beginning. 
 

Remember

: If unipolar electrodes are used the whole slice will be stimulated resulting in a DC 

shift of the whole slice. 
 

4.2. FILTERED/UNFILTERED  switch (rear panel) 

 
There  is  a  switch  at  the  rear  panel  of  the  ISO-STIM  01D  labeled  ISOLATED  OUTPUT 
GROUND. In most configurations of the setup switch to FILTERED can reduce the noise, but 
this  depends  on  several  factors,  especially  grounding  of  the  whole  setup.  Therefore,  it  is 
unpredictable whether switching from UNFILTERED to FILTERED substantially improves 
the signal-to-noise ratio. 
 

Summary of Contents for ISO-STIM 01D

Page 1: ...FOR THE ISO STIM 01D STIMULUS ISOLATION UNIT 100 V 10 mA bipolar output VERSION 4 3 npi 2016 npi electronic GmbH Bauhofring 16 D 71732 Tamm Germany Phone 49 0 7141 9730230 Fax 49 0 7141 9730240 suppor...

Page 2: ...nts 4 3 ISO STIM 01D Stimulus Isolation Unit 5 3 1 System Description 5 3 2 Description of the Front Panel 6 4 Operation 9 4 1 Connections Noise Reduction 9 4 2 FILTERED UNFILTERED switch rear panel 9...

Page 3: ...re connecting any system to mains Always use a three wire line cord and a mains power plug with a protection contact connected to ground protective earth Before opening the cabinet unplug the instrume...

Page 4: ...ON Observe extreme caution when working with this instrument 1 Always connect high voltage power supplies to protective earth 2 Do not touch connections unless the instrument is turned off and the cap...

Page 5: ...mplitude and polarity The ISO STIM 01D has three modes of operation In the DIRECT input mode the output signal follows exactly the input signal In the GATE TTL input mode the output signal is generate...

Page 6: ...URATION potentiometer Potentiometer that sets the duration of the stimulus multiplied with the duration range factor generated by the built in timing unit ANALOG INPUT MODE GATE TTL MODE The duration...

Page 7: ...OUTPUT MODE switch set to CURRENT The polarity of the output signal is set by switch 7 Important If the CAP COMP 15 is overcompensated in CURRENT MODE the stimulus isolator will ring and not work pro...

Page 8: ...VOLTAGE trim pot Trim pot to compensate for the VOLTAGE OFFSET of the stimulating electrode in TTL mode It is recommended to compensate the offsets only in a completely warmed up condition i e after...

Page 9: ...possible source of noise is the chassis of the stimulator In a noisy environment it may act as an antenna picking up 50 Hz or 60 Hz noise respectively especially when operating with current stimuli In...

Page 10: ...DE VOLTAGE i e an input of 1 V will lead to an output stimulus of 10 V In CURRENT OUTPUT MODE the amplitude of the output stimulus is again dependent on the setting the CURRENT RANGE switch With the C...

Page 11: ...voltage divider between the electrode resistance and the protection resistor 4 4 ISO CAB optional accessory npi provides an optional shielded cable for connection of stimulation electrodes One end is...

Page 12: ...__________________________________________________________________________________________________ ___________________________________________________________________________ version 4 3 page 12 Figur...

Page 13: ...osition CURRENT CURRENT RANGE switch set to x1 A and the amplitude of the input signal is 2 V The max signal at the isolated output plugs has to be 20 V according to the current of 200 A flowing throu...

Page 14: ...with 3 position rotary switch 100 V voltage source Time range factor x10 s x100 s x1 ms selectable with toggle switch time range up to 99 ms Amplitude potentiometer 3 digits 10 V amplified internally...

Reviews: