Installation / Operation / Maintenance 

Sentinel Benchtop Ionizing Blower 

Model 6210 

Overview 

 

The Sentinel Benchtop Ionizer utilizes DC high voltage to 
generate balanced ionized airflow to eliminate static charges 
on surfaces.  The Ionizer can be placed at one end of the 
workbench, wall or shelf mounted, and directed at the target 
or area to be neutralized.  The target’s neutralization time 
will be optimal when placed 12” to 48” directly in front of 
the Ionizer and will increase as the distance increases. 

 
 

General Guidelines 

 

 

Keep water, oil, contamination and any fluids away 
from the Ionizer at all times. 

 

Keep the Ionizer clean inside and out for optimal per-
formance. 

 
 

CAUTION: DISCONNECT THE IONIZER FROM 
POWER BEFORE PERFORMING ANY MAINTE-
NANCE, INSPECTION OR CLEANING TO AVOID 
AN ELECTRICAL SHOCK HAZARD AND TO PRE-
VENT DAMAGE TO THE IONIZER. 

 
 

Contents 

 

Remove the Ionizer from the carton and inspect for damage.  
Included with the Ionizer should be: 

 

 

One (1) Sentinel Benchtop Ionizing Blower with Stand 
and adjustment knobs installed.

 

 

One (1) Detachable IEC Power cord (US 115V or 230V, 
UK or Euro; must be specified at time of order).

 

 

One (1) Round foam air filter with (1) 

¾

” round hook 

fastener dot.

 

 

This Installation, operation, and maintenance instruc-
tions sheet.

 

 
 

Installation 

 

Locate the Ionizer 12” to 48” from the target to be neutral-
ized for optimal performance, making sure that the airflow 
will not be restricted.  The Ionizer comes with a stand so that 
it can be positioned on a flat surface, aimed at the target and 
locked via adjustment knobs.  It can also be wall or 

 

 

 
 

shelf mounted to keep valuable bench space free for other 
purposes and equipment.  A working electrical outlet within 
reach of the 6’ long detachable cord that is supplied with the 
Ionizer will be required.  The outlet must be a properly in-
stalled and grounded 3-prong outlet with supplied voltage to 
match the Ionizer being installed.  If one is not present, con-
sider relocating the Ionizer to be within reach of an outlet or 
having an electrical outlet installed near the Ionizer.  Be sure 
the ON/OFF switch, located on the front of the Ionizer, is in 
the “OFF” position.  Plug the power cord into the back of the 
Ionizer and then into the appropriate AC power source. 

 
 

IMPORTANT NOTE: 

Before powering up the Ionizer, 

verify that the AC outlet is properly grounded and that the 
supplied line voltage matches the electrical specification on 
the label at the bottom of the Sentinel Benchtop Ionizer.  
Plugging the Ionizer into an outlet with the incorrect voltage 
may damage the Ionizer and will void any warranty cover-
age.  The Ionizer must have a good ground to maintain 
proper balance.

 

 

Summary of Contents for Sentinel 6210

Page 1: ...on and maintenance instruc tions sheet Installation Locate the Ionizer 12 to 48 from the target to be neutral ized for optimal performance making sure that the airflow will not be restricted The Ionizer comes with a stand so that it can be positioned on a flat surface aimed at the target and locked via adjustment knobs It can also be wall or shelf mounted to keep valuable bench space free for othe...

Page 2: ... timum neutralization efficiency and operation cleaning should be performed on a regular basis once a month is rec ommended In addition to improving decay time cleaning also keeps the Ionizer balanced without the need for adjust ment NOTE The AC power cord MUST be disconnected before the Ionizer can be disassembled for maintenance First turn the Ionizer OFF and unplug the power cord Remove the bac...

Page 3: ...or a Return Authorization number and ship ping address IMPORTANT NOTE In some cases if the balance sensing grill inside the blower becomes contaminated the Ionizer may not be able to detect an out of balance condition and may continue to run in a green light status The presence of contamination on the backside the side nearest the fan of the balance sensing grill can cause the Ionizer to overcompe...

Page 4: ... Air Velocity Two speed fan 250 lfpm with fan on high speed Balance 3 volt typical Temperature range 65 F to 80 F Relative humidity 15 to 65 Ion Emission Steady state DC with sensor feedback Emitter Pin Electrodes 050 diameter 316 stainless steel 8 required High Voltage Power Supply 5 5 kV DC nominal 1 positive HVDC and 1 negative HVDC Input Power AC line power internally selectable for 100 120 VA...