STANDARD ACCESSORIES
To make the warranty active it is necessary that a copy of the supplied Unit Passport is sent, through
the agent, to the manufacturer; for want of this the warranty will decline
.
2.7
TECHNICAL SPECIFICATIONS
2.7.1
Environment operating conditions
The sterilizer is designed to operate in environments with temperature between 3 °C and 40 °C, relative hu-
midity <95%, atmospheric pressure from 750 mBar to 1050 mBar and altitude between 0 and +2000m.
Chamber dimensions
Ø = 240 mm
Depth = 384 mm
Auto-switching-off
elapsed 30’ from the end of a
cycle and without any action
Chamber capacity
17,5 l
Dual water tank
4 liters each (waste and
fresh water tanks)
Maximum load
4 kg (solid)
1,5 kg (porous)
Bacterial filter
0.3 µm al 99.97 %
Warming-up time
20’ from room temperature
10’ with pre-heated chamber
Clock battery
(may be replaced only by au-
thorized service)
Varta CR2032
Sterilization time
from 3’ to 90’ depending on the
selected cycle
Transmitted heat
in environment
0,22 J/h at 23°C
Drying time
from 6’ to 14’ depending on the
selected cycle
Sound emission
52 dB/A at 1 m
External dimensions
443 x 590 x 428 mm
(L x D x H)
Working cycle
continuous
Net weight
47 Kg
Pollution grade
2
Power supply voltage
230 Vac
Transient overvoltage
II
Frequency
50 / 60 Hz
Water conductivity
monitoring
yes
Max consumption
1920 W
Maximum volume available
on the trays
10 l
Average consumption
1000 W
Maximum chamber
temperature
135°C (-0/+2°C)
Standby consumption
10 W
Safety valve
working pressure
2,55 bar
AC Fuses
2 x 12,5A T (type 6.3 x 32 CT ) -
IEC 127
NEVER USE IN PRESENCE OF FLAMMABILE OR ANAESTHETIC GAS.
Tray rack (P/N 105078)
Operating Manual
Warranty paper
Cleaning tabs
Bacterial filter
(P/N 021008)
(+ 1 pcs extra)
Water filter
(P/N 105320)
(+ 2 pcs extra)
Filter spanner
(P/N 105228)
2+2 Trays
(P/N 105076 - 105077)
Tray holder
(PN/ 105619)
Drain tube
(P/N 119001)