35
Nucleus Freedom Speech Processor
Do not let children replace batteries without adult
supervision.
Make sure that batteries are kept out of reach of
young children.
Batteries can be harmful if swallowed. If swallowed,
seek prompt medical attention at the nearest
emergency center or Poisons Information Center.
BTE Controller
Your Freedom BTE Controller uses three high power
675 zinc air batteries. Silver Oxide or alkaline batteries can
also be used in very humid conditions: your Freedom BTE
speech processor will only operate for a few hours when
Silver Oxide or alkaline batteries are used. Do NOT use
rechargeable NiMH button cells (HB 116/054).
High Power 675 zinc air batteries are sealed, usually with a
tab. Always remove the seal or tab at least one minute before
using, to allow air access to the battery, so it can operate.
To change the batteries:
1.
Remove the new set
of batteries from their
pack and make sure
the seals are removed.
2.
Let the new batteries
stand for one
minute, outside of
their packaging with
their seals removed,
to allow the air to
activate the batteries.
Seal
Battery
3.
Turn off the speech processor and take it off.
N31018F Iss1D03textmas.indd 35
N31018F Iss1D03textmas.indd 35
24/05/2005 4:14:59 PM
24/05/2005 4:14:59 PM