able to change the mode, intensity etc. It will avoid screen touching by mistake. To unlock the
screen, you can press the ON/OFF once slightly or press the lock icon on the screen.

Pressing the countdown time on the screen, you can set the timer from 10 minutes to 60 minutes.

When power bar in the battery icon disappear, the unit is going to run out power, it should be
charged as soon.

When you turn on the unit, if the unit test no load, for example you didn’t connect the unit well
with body, you are not able to adjust the intensity.

Recommended practice:

Duration suggested for each skin area is 20 min and 2 times per day. Consult with your physician
for longer and more frequent uses.

Start from the lowest intensity and gradually adjust the intensity to a comfortable level at a scale
from 1 to 20.

Good skin care is important for a comfortable use of device. Be sure the treatment site is clean of
dirt and body lotion.

Keeping the electrode in the storage bag after use will extend its lifespan. The electrode is
disposable and should be replaced when it loses the adhesiveness. To purchase additional electrodes,
please contact the seller.

Summary of Contents for PL-029U

Page 1: ...Electronic Pulse Stimulator TENS Unit Operating Manual PL 029U Edition V1 0...

Page 2: ...ards 5 Electromagnetic Compatibility and FCC Compliance statement 6 Product Specifications 12 Setup 15 Operating Instruction 16 Recommended practice 17 Cleaning and Maintenance 18 Trouble Shooting 19...

Page 3: ...Contents of Different Language Versions English 1 21 French 22 36 German 27 30 Italian 31 35 Spanish 36 40...

Page 4: ...ntial coming from the central nervous system to trigger contraction of the muscle It may be helpful in relieving aches and pains in various parts of the body such as the waist shoulders joints hands a...

Page 5: ...e Acute disease malignant tumor infective disease pregnant heart disease high fever abnormal blood pressure lack of skin sensation or an abnormal skin condition any condition requiring the active supe...

Page 6: ...electrical stimulation or electrical conductive medium gel 8 if the user has suspected or diagnosed epilepsy the user should follow precautions recommended by his or her physician 9 to use caution if...

Page 7: ...timulation electrodes applied to the skin Patients may experience headache and other painful sensations during or following the application of electrical stimulation near the eyes and to the head and...

Page 8: ...tial performance Collateral standard Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment EN 980 Symbols for use in the labeling of medi...

Page 9: ...orrect operation of the unit 3 Caution This unit has been thoroughly tested and inspected to assure proper performance and operation 4 Caution This machine should not be used adjacent to or stacked wi...

Page 10: ...internal battery powered Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth Not applicable internal batt...

Page 11: ...nity The device is intended for use in the electromagnetic environment specified below The customer or the user of the device should assure that it is used in such an environment Immunity test IEC 606...

Page 12: ...dcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field s...

Page 13: ...bove the recommended separation distance din metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter i...

Page 14: ...in a particular installation If the product does cause harmful interference to radio or television reception which can be determined by turning the product on or off the user is encouraged to try to c...

Page 15: ...us Microseconds Type of protection against electric shock Internally powered equipment Pulse frequency 1 100Hz Hz vibration per second Grade of waterproof IP22 Output Voltage Max 50pp 20 at 500ohm loa...

Page 16: ...ull capacity after fully charged The time required for me equipment to cool from the maximum storage temperature between uses until it is ready for intended use 15 minutes Typical service life of Batt...

Page 17: ...30 40 50 60 1 100 100 Mode 7 10 20 30 40 50 60 1 35 100 Mode 8 10 20 30 40 50 60 6 6 10 100 Mode 9 10 20 30 40 50 60 4 8 100 Mode 10 10 20 30 40 50 60 62 5 100 Mode 11 10 20 30 40 50 60 2 100 Mode 12...

Page 18: ...e connect unit onto the electrode pad ON OFF Screen lock Intensity Increase Countdown Timer Battery Indicator LCD Display Intensity Decrease Intensity Indication A B Channel Output A Channel USB Port...

Page 19: ...icon A will be flashing you can change the intensity of channel A now If you want to change the intensity of channel B press the button A B once or you can just simply press A or B icon then the icon...

Page 20: ...e unit well with body you are not able to adjust the intensity Recommended practice Duration suggested for each skin area is 20 min and 2 times per day Consult with your physician for longer and more...

Page 21: ...se use wipe to clean the device first and then use the dry cloth to wipe it again The electrodes coming with the device are disposable and should be replaced when their adhesiveness becomes worse Cont...

Page 22: ...ed The adhesive surface of the pads is dirty or dry Wash adhesive surface of pads gently with your fingertips for about 3 seconds under slow running water The therapy time is too long or the intensity...

Page 23: ...20 Recommended Use Positions...

Page 24: ...21 Contact Information JKH Health Co Ltd Address 4 5F Building 12 Hengmingzhu Industrial Park Xinqiao Tongfuyu Industrial Area Shajing Baoan District Shenzhen China...

Page 25: ...22 lectrostimulateur Appareil TENS Installation D ballez le produit de son emballage retirez le produit et les accessoires et branchez les lectrodes l appareil...

Page 26: ...uement modifier l intensit du canal A Si l ic ne B clignote vous pouvez uniquement modifier l intensit du canal B Si vous souhaitez passer du canal A au canal B vous pouvez appuyer sur le bouton A B u...

Page 27: ...ous pouvez appuyer une fois l g rement sur le bouton ON OFF ou appuyer sur l ic ne de verrouillage l cran En appuyant sur la minuterie l cran vous pouvez r gler la minuterie de 10 minutes 60 minutes L...

Page 28: ...dh rent pas fermement la peau Appliquez les lectrodes de sorte qu elles adh rent fermement la peau Vous n avez pas retir le film de protection des lectrodes Retirez le film de protection des lectrodes...

Page 29: ...rotection des lectrodes Vous avez appliqu les lectrodes imm diatement apr s les avoir nettoy es Faites s cher les lectrodes La face adh sive des lectrodes est ab m e Remplacez les lectrodes La face ad...

Page 30: ...27 Elektronischer Pulsstimulator TENS Ger t Einrichtung ffnen Sie die Verpackung des Produkts entnehmen Sie es und alle Accessoires Verbinden Sie die Elektroden Pads mit dem Ger t...

Page 31: ...Symbol an den zwei Intensit tsleisten am unteren Bildschirmrand Wenn ein bestimmter Modus ausgew hlt wurde k nnen Sie die Intensit t dieses Modus anpassen Wenn das Symbol A blinkt k nnen Sie die Inten...

Page 32: ...e Intensit t nicht anpassen Empfohlene Anwendung Die empfohlene Anwendungsdauer pro Hautbereich betr gt 20 Minuten zweimal pro Tag Besprechen Sie dies auch mit Ihrem Arzt f r l ngere Verwendungen Begi...

Page 33: ...unden unter leicht flie endem Wasser Die Behandlungszeit ist zu lang oder die Intensit t zu hoch Verringern Sie die Anwendungszeit oder die Intensit t Die Elektro Pad Oberfl che ist verbraucht Elektro...

Page 34: ...31 Stimolatore Elettronico a Impulsi Unit TENS Layout Aprire la confezione estrarre il prodotto e gli accessori e collegare gli elettrodi al dispositivo...

Page 35: ...icemente premere l icona A o quella B sui due lati delle barre di intensit nella parte inferiore dello schermo Una volta scelto un determinato Modo possibile modificarne l intensit Se l icona A lampeg...

Page 36: ...cona della batteria scompare l unit sta per restare priva di alimentazione e la batteria deve essere ricaricata non appena possibile Quando si accende l unit e non viene rilevato alcun carico ad esemp...

Page 37: ...atteria La pelle si arrossa o si irrita La superficie adesiva delle pastiglie sporca o secca Lavare la superficie adesiva delle pastiglie delicatamente con i polpastrelli per circa 3 secondi sotto l a...

Page 38: ...la seguente i problemi usuali e le soluzioni suggerite Se l azione consigliata non risolve il problema si prega di contattare il venditore Le pastiglie sono state conservate a temperatura elevata espo...

Page 39: ...36 Estimulador el ctrico transcut neo Unidad de TENS Configuraciones Saque el producto y los accesorios del paquete y encaje la unidad de conexi n en la almohadilla del electrodo...

Page 40: ...arte inferior de la pantalla para cambiar el canal Cuando haya seleccionado cierto modo cambie la intensidad del modo El icono A parpadear y podr cambiar la intensidad del canal A Si desea cambiar la...

Page 41: ...ezca la unidad se descargar y deber cargarla pronto Cuando encienda la unidad si sta no carga por ejemplo que no haya conectado bien la unidad con el cuerpo no podr ajustar la intensidad Uso recomenda...

Page 42: ...parentes todav a est n pegadas a las almohadillas Quite la pel cula sobre la superficie adhesiva de las almohadillas Las almohadillas est n pegadas o se superponen No junte las almohadillas o sobrepon...

Page 43: ...tamente despu s del lavado Seque la almohadilla Est da ada la superficie adhesiva de la almohadilla Vuelva a colocar la almohadilla La superficie adhesiva de la almohadilla no es pegajosa Las almohadi...