Voyager IV
Nitrox System
Nuvair www.nuvair.com
Page
21
14 Pre-Operation of Compressor Before Producing Nitrox
14.1 Calibrate Oxygen Analyzers
Gas production is monitored with the Permeate Oxygen Analyzer before entering the HP Compressor
to obtain a rough estimate of %O
2
(+/- 2%); however, do not rely on this reading as an indication of
%O
2
at the HP Compressor outlet. Prior to pumping Nitrox from the Compressor, it must be
monitored with the Fill Oxygen Analyzer to obtain a precise measurement of %O
2
(+/- 1%).
Both
Oxygen Analyzers must be calibrated prior to each use.
Oxygen Analyzers must be calibrated before each use. See Oxygen Analyzer manuals for
correct calibration procedures. Improper calibration of the Fill Oxygen Analyzer may result in
the use of incorrect Nitrox mixtures, which may cause serious injury or death to the diver
using the gas mixture.
At altitudes above sea level, a correction factor must be used when calibrating the Fill Oxygen
Analyzer may not be achievable. See Fill Oxygen Analyzer manual for correcting analyzer
readings at various altitudes. Improper calibration of the Fill Oxygen Analyzer may result in
the use of incorrect Nitrox mixtures, which may cause serious injury or death to the diver
using the gas mixture.
The Permeate Oxygen Analyzer supplies oxygen readings that can vary +/- 2% O
2
due to heat,
humidity, and pressure changes experienced in the Nitrox flow and therefore should only be
used for rough estimates of %O
2
. The Fill Oxygen Analyzer supplies more accurate oxygen
readings, /- 1% O
2
. For Scuba cylinder Nitrox fills, the user must always verify the final
fill with a third independent Oxygen Analyzer.