86
9.6 Adjust
Warning
:
The following functions must be carried out under the
guidance of professionals, if the operation is wrong, it may lead to
the instrument can’t work properly!
Press [MENU]→[F4] →[1](Adjust) to enter the interface of
―Adjust‖:
Adjust
1. Calibrate I.E
2. Calibrate TILT:X
3. Calibrate TILT:Y
9.6.1 Calibrate I.E
To adjust index error, the system will first ask you to ‗Aim at a
target F1‘(face left). You may press [ENT] to confirm after aiming,
and the instrument will measure its vertical angle automatically. The
angle will be displayed for one second and the system will ask you to
‗Aim at a target F2‘(face right). Following the same steps, the
instrument will then calculate the index error and display it.
Now press [ENT] to save the index error (calibration done),
otherwise press [ESC] to quit (previous I.E remains).
Note:
If there is no special requirement, the compensator should
be turned on before Index error correction.
9.6.2 Calibrate TILT:X
Before compensating for the compensator, make sure that the
indicator difference is recalibrated in accordance with 9.6.1 procedure
in the closed compensator state.
When calibrating tilt, please make sure that the tilt is off and the
index error is little. If you want to correct the index error, please refer
to the calibrate I.E step.
First,
p
lace the instrument as picture shown below with
collimator facing up. This will help screw A to adjust the inclination
of the instrument.