Fiber waste collector bin counter limit
Threshold level for maximum allowed number of
cleaves made before the warning “Waste bin limit reached” is displayed on the
controller software and the waste collector bin should be emptied. If limit is set to zero (0), the counter function is disabled. When the
collector bin is emptied, reset the co
unter with the function “Empty fiber waste collector bin” under
the Service menu.
Fiber waste vacuum on-time
A vacuum on delay time of the fiber waste removal system. Increase time if waste fiber tends not to be removed after cleaving (this
will increase air consumption of the AUTOCLEAVER
TM
). By setting time to zero (0) the exhaust system will be deactivated (not
recommended).
Touch-start threshold level
Required force to start the cleaving cycle on an AUTOCLEAVER
TM
by pressing on the left side on the rail or adaptor plate towards the
right. If this value is set to zero (0) the feature is disabled.
Touch-confirm threshold level
Required force for confirming any messages etc. on an AUTOCLEAVER
TM
. If this value is set to zero (0) this feature is disabled.
Maximum movement to reach set tension
This parameter specifies the maximum allowed distance for the tensioning motor to move from the clamping position to reach the
desired cleaving tension. If this distance is reached before the desired tension is reached the error message "Fiber slipping?" or
alternatively "No fiber?" will be displayed.