34
e500 User Manual
(15PL1011 - Rev.7 6_6_2017) English version
Table-5 Ventilation Alarms
Symbol Name
Priority
Visual Alarm
Audible
Alarm
Alarm symbol
Warning LED
Patient Circuit
disconnect (Breathing
circuit Integrity)
High after 15
seconds
Flashing
symbol
Flashing Yellow for
15 seconds then
Red
15 sec. delay Two
bursts with five
pulses
Low Airway Pressure
High
Flashing
symbol
Red
Two bursts with
five pulses
High Airway Pressure
High
Flashing
symbol
Red
Two bursts with
five pulses
Leak (at 40%
below set Vt)
High
Flashing
symbol
Red
Two bursts with
five pulses
No Oxygen
≤
20 PSI
High
Flashing
symbol
Red
Two bursts with
five pulses
Low Oxygen
≤
40 PSI
Medium
Flashing
symbol
Yellow
One burst with
three pulses
High input pressure
≥
90 PSI
High
Flashing
symbol
Red
Two bursts with
five pulses
APNEA High
Flashing
symbol
Red
Two bursts with
five pulses
Empty Battery
High
Flashing
symbol
Red
Two bursts with
five pulses
Low Battery
Low
Flashing
symbol
N/A N/A
Pause N/A
Flashing
symbol
Yellow every 15
seconds
N/A
Play N/A
Flashing
symbol
N/A N/A
Lock N/A
Solid
symbol
N/A N/A
Alarm Silence
N/A
Solid symbol
N/A
N/A
Patient effort
Low
Solid symbol
during Patient
effort
N/A N/A
Invalid setting- Refer
to manual
N/A
Solid symbol
During invalid
selection
N/A N/A
Setting Conflict
N/A
Solid symbol
During invalid
selection
N/A N/A
Confirm
N/A
Flash symbol
after primary
selection
N/A N/A