39
e500 User Manual
(15PL1011 - Rev.7 6_6_2017) English version
Preventive Maintenance
To ensure proper operation of the ventilator, regular inspection and checking
of the device and accessories for correct function should be undertaken by a
responsible member of staff. It is recommended that a preventive maintenance
record be maintained for each unit.
The battery should be charged and the ventilator be checked for leakage and
proper function at least every six months, and more frequently in high use
applications. Any malfunction unit should be returned to the manufacturer or
an authorized service center since this product is not designed for field
disassembly or service. Unauthorized repairs will nullify the product warranty.
Level II service
The device shall be returned to the manufacturer or a service center
authorized by the manufacturer for bi-annual Level II service every 2 years.
Manufacturer full service
The device shall be returned to O-Two Medical Technologies for Manufacturer
Full Service every 6 years.
Warning
Any malfunction unit should be returned to the manufacturer or an authorized
service center since this product is not designed for field disassembly or
service.