47
e500 User Manual
(15PL1011 - Rev.7 6_6_2017) English version
9.6 Oxygen Consumption
For a “D” size cylinder (capacity of 425 liters), pressurized to 2015 PSI and
with the e500 set to 500/10, 100% Oxygen) the pneumatic operating time is
85 minutes without PEEP (
0.2 min/L)
and is 39 minutes with maximum PEEP
(0.089 min/L).
Duration of consumption of a cylinder in minutes is calculated approximately
by dividing the liter content by the minute volume with or without PEEP.
9.7 Oxygen Concentration delivered against different back
pressures
0
10
20
30
40
50
60
70
80
90
100
0
20
40
60
80
100
120
Flow Rate
L/min
O
2
Concentration
%
0 cmH2O
60 cmH2O
30 cmH2O
5 cmH2O
15 cmH2O