6
e500 User Manual
(15PL1011 - Rev.7 6_6_2017) English version
3.2 Function Keys
While all ventilation parameter settings are controlled by the Control Selection
Knob (M) in Figure 1, there are a number of key membrane buttons which
control additional ventilator functions:
3.2.1 ON/OFF
To turn on the ventilator, Press the ON/OFF button (N) in Figure 1 for one
second, during that second the associated green LED will start flashing at a
high frequency. After 1 second the ventilator will turn on but without
ventilation until the appropriate
tidal volume/frequency combination is
selected from the dial. If the button is pressed and released for less than a
second, the ventilator will remain OFF.
To turn off the ventilator, Press and hold the ON/OFF button for 4 seconds,
the green power LED will start flashing at a high frequency. After 4 seconds
the ventilator will turn off and all ventilation will stop. If the button is pressed
and held for less than 4 seconds, the ventilator will stay ON.
Warning
At approximately 2% of full battery capacity, the ventilator will not start
when in the off position or will shut down when operating.
3.2.2 Control Selection Knob
The Control Selection Knob (M) in Figure 1 is used to navigate between
parameters, change modes, select primary function change when rotated,
and to confirm function changes when pressed.