Specifications
Temperature range
35629-00
-32 to 400°C (-25 to 750°F)
35629-10
-32 to 545°C (-25 to 950°F)
D : S
12 : 1 (35629-10 has 90% encircled energy at the
focal point)
Display Resolution
0.2°C (0.5°F)
Accuracy
For targets:
(assumes ambient
Above 23°C (73°F): ±1% of reading or ±1°C (± 2°F),
operating temperature
whichever is greater
of 23-25°C[73-77°F])
-18 to 23°C (0 to 73°F): ± 2°C (± 3°F)
@ calibration geometry*
-26 to -18°C (-15 to 0°F): ± 2.5°C (± 4°F)
-32 to -26°C (-25 to -15°F): ± 3°C (± 5°F)
Repeatability
± 1°C (± 2°F), or ± 0.5% of reading, whichever is greater
Response time
500 mSec
Spectral response
8–14 µm
Emissivity
pre-set 0.95
Ambient operating range
0 to 50°C (32 to 120°F)
Laser Max
Laser turns off above 40°C (104°F) ambient temperature
Relative humidity
10–95% RH noncondensing, at 30 to 50°C (86 to 122°F)
Storage temperature
-20 to 60°C (-13 to 158°F) without battery
Weight / Dimensions
320g (11oz); 200 x 160 x 55 mm (8 X 6 X 2 in)
Power
9V Alkaline or NiCd battery
Typical Battery life
10 hours with laser and backlight on
(Alkaline)
40 hours with laser and backlight off
Tripod Mount
1/4” 20 UNC threading
*Unit calibrator is 279.4mm (11 in.) from the 140mm (5.5 in.) 0.95 emissivity backbody.