10.  Specifications

Mode

pH

Temperature

Conductivity

0 to 19.99 µS,

Range

0.00 to 14.00 pH

0.0 to 100.0˚C

0 to 199.9 µS,
0 to 1999 µS,
0 to 19.99 mS

0.01 µS, 

Resolution

0.01 pH

0.1˚C

0.1 µS, 
1.0 µS, 

0.01 mS

Accuracy

±0.01 pH

±0.5˚C

±1% full scale or ±1 digit

Up to three points

Calibration

(pH 4.01, 7.00, 10.01)

Offset in 0.1°C

Up to four points

with automatic 

increments

(one point per range)

buffer recognition

Temperature compensation:

automatic from 0 to 50°C

Conductivity cell constant (K):

1.0

Conductivity temperature coefficient:

2.00% per °C

Operating temperature: 

0 to 50°C

Power:

four 1.5 V AAA batteries (included) or AC adapter (optional; order 

separately on page 22)

Battery life: 

> 50 hours

Dimensions:
35630-00
Meter: 7.5"L x 3.5"W x 1.75"H (19.1 cm x 8.9 cm x 4.5 cm)
Boxed: 9.2"L x 8.5"W x 2.75"H  (23.3 cm x 21.6 cm x 7 cm)
Probe: 6.8"L x 1.3" dia (173 mm L x 32 mm dia)

35630-02
Meter: 7.5"L x 3.5"W x 1.75"H (19.1 cm x 8.9 cm x 4.5 cm)
Boxed: 9.2"L x 9.2"W x 2.75"H  (23 cm x 23 cm x 7 cm)
Probe: 6.8"L x 1.3" dia (17.3 cm L x 3.2 cm dia)

Shipping weight: 

1.4 lbs (0.64 kg)

20

9.  Error Messages

LCD

Indicates

Cause

Solution

Display

Err

Unrecognized input

Wrong input in

Release key.  Select 

annunciator

from keypad

selected mode.

valid operations
depending on mode.

CAL & Err

Calibration error

Wrong value input 

Check your input value, 

annunciators

at calibration.

clean probe.

blink

Dirty probe. 

See Calibration sections or 
Probe Maintenance section.

Battery 

Low battery level

Need new batteries

Clean battery contacts.

indicator blinks

or battery 

Replace batteries with fresh

connection is bad

ones, noting polarity

Err. 1

Memory write

Instrument too

Turn meter on and off

(in primary

error

old (>10 years).

again. If message persists, 

display)

Hardware failure

return unit*

Err. 2

Checksum

Batteries too weak.

Press ENTER, then turn 

(in primary

error

Hardware failure.

off meter. Change

display)

batteries. Recalibrate
Return*

Err. 3

A/D converter error

Faulty hardware

Return*

* See "Warranty" and "Return of Items" on page 24

If an error message appears in the primary display (the upper row of larger digits),
switching off the meter and switching it on again may eliminate the error message. 

See figure below.

If error persists, or the meter shows incorrect values, return the meter.  

For a complete diagram of the display, see page 4. 

ERR 1 in primary display

Summary of Contents for 35630-00

Page 1: ...will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Au...

Page 2: ...ting 19 9 Error Messages 20 10 Specifications 21 11 Accessories 22 23 12 Warranty 24 13 Return of Items 24 2 3 pH Conductivity C Meter HOLD MODE ON OFF CAL MEAS C MEAS pH ATC pH Con 10 Series ENTER 1...

Page 3: ...nding graphic indicator on the LCD ON OFF Powers and shuts off the meter HOLD Freezes the measured reading To activate press HOLD while in measurement mode To release press HOLD again MODE Selects the...

Page 4: ...s a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes or electrodes For a replacement probe see the Accessories section page 22 NOTE Keep connector dry and c...

Page 5: ...bration See section 4 5 on pages 14 15 for temperature calibration Preparing the Meter for Calibration The pH conductivity temperature probe included with this meter is designed for use with this mete...

Page 6: ...ate the pH standard buffer solution See figure NOTE If using a pH buffer other than pH 7 press the or keys to scroll up or down until the secondary display value is the same as your pH buffer value pH...

Page 7: ...containers 2 If necessary press the MODE key to select Conductivity Mode The S or mS indicator will appear on the right side of the display 3 Rinse your probe with deionized water then rinse the probe...

Page 8: ...and the secondary display shows you what the temperature value was initially See figure 4 5 ATC CAL Temp C B A B pH ATC CAL pH pH A 4 Compare the primary display reading to a NIST traceable thermomete...

Page 9: ...electrode has dehydrated soak it for 30 minutes in OAKTON electrode storage solution or a 2M 4M KCl solution 2 Switch on the meter The MEAS annun ciator appears on the top center of the LCD The ATC i...

Page 10: ...a Clean probe See Probe Care Maintenance p 18 18 7 Probe Care and Maintenance Under typical operating conditions the probe will need to be replaced every 6 to 12 months In extreme applications the pr...

Page 11: ...Error Messages LCD Indicates Cause Solution Display Err Unrecognized input Wrong input in Release key Select annunciator from keypad selected mode valid operations depending on mode CAL Err Calibratio...

Page 12: ...D 00653 04 Electrode storage solution 1 pint bottle Keeps electrode bulb moist for faster more accurate readings Shpg wt 1 1 lbs 510 gms WD 00653 06 Electrode cleaning solution 1 pint bottle Removes b...

Reviews: