Printed in the U.S.A. 8/01 OAKTON TM Reg. #1,692,543. We reserve the right to make changes, improvements and modifications to products shown.
All prices subject to change without notice. OAKTON Instruments are available only through authorized OAKTON distributors.
General Applications:
Ideal for any application that requires measuring/monitoring the temperature of any liquid, solid, semisolid or gel.
Food Processing/Agricultural Applications:
Use as a practical substitute for glass thermometers where the danger of product
contamination exists. Use with a penetration probe for insertion into semisolids like fruits, semifrozen foods or soil.
Industrial Applications:
Use in photo developing, chemical and plating industries, and all applications requiring temperature measurement.
Laboratory Applications:
Monitor heating or cooling samples and reagents.
Educational Applications:
An ideal student thermometer—select °C or °F with the push of a button, and avoid dangers from
broken glass or mercury spillage.
HVAC Applications:
Use Min/Max feature to monitor heating and cooling efficiency, flue gas temperatures, boiler water, heater jacket
temperatures, inlet and outlet water heater temperatures, etc.
• ORDER FROM •
Optional rubber boot
offers sturdy
protection for meter and acts as meter
stand for benchtop use.
Large Custom LCD
for
clear and easy readings.
Splash-resistant membrane keypad
helps protect components.
•
Applications
•
•
Specifications
•
•
Ordering Information
•
Easy access connection
for
thermocouple miniconnector.
ACORN
™
Temp TC thermocouple meters
ACORN
™
Temp TC Meters
All meters include operating instructions and batteries. Temperature probe not included, order separately. Shipping weight: 1.0 lb (0.5 kg).
WD-35627-00 Temp TC meter
WD-35627-02 Temp TC meter with protective boot
WD-35627-05 Temp TC meter
with general purpose type K thermocouple probe (WD-08439-62), 4.5"L with 5-ft coiled cable.
Temp range: –418 to 1650°F (–250 to 899°C). Response time: 30 seconds (in liquids)
WD-35627-80 Protective rubber boot for Temp TC meter
, encases meter in sturdy rubber to protect it from drops and dings.
Also features meter stand for convenient tabletop use. Shipping weight: 0.5 lbs (0.2 kg)
WD-09376-00 Replacement batteries
, 1.5 V AAA. 12/pack
Bulletin TEMP B3 R1
Calibration:
factory calibrated, push-button offset adjustment
Input connector:
standard miniconnector
Display:
single line LCD
Operating Temperature:
0 to 50°C
Power:
4 x 1.5 V AAA batteries >200 hours continuous use (included)
Probe:
use any type J, K or T thermocouple probe with mini-connector
(not included)
Dimensions:
Meter: 5.5" x 2.7" x 1.3" (14 x 7 x 3.5 cm)
Boxed: 9.25” x 6.5" x 3.0" (23.5 x 16.5 x 7.5 cm)
Weight:
Meter only: 0.9 lb (0.4 kg); Boxed: 1 lb (0.5 kg)
Type J: –200 to 1000°C/–328 to 1832°F;
Range
Type K: –250 to 1372°C/–418 to 2501°F;
Type T: –250 to 400°C/–418 to 752°F
Resolution
0.1° from –99.9°C to 299.9°C / °F
1° outside this range
±0.25% of reading plus 1°C / 2°F
Accuracy
below –99.9°C / °F,
(meter only)
±0.25% of reading plus 0.5°C / 0.9°F
above –99.9°C / °F
Catalog Number
Probe type
Temperature range
Response time
General purpose probes.
WD-08517-55
J
–90 to 760°C / –310 to 1400°F
15 seconds (in liquids)
WD-08516-55
K
–250 to 899°C / –418 to 1500°F
15 seconds (in liquids)
WD-08500-55
T
–250 to 400°C / –418 to 752°F
15 seconds (in liquids)
Penetration probes.
Use for penetrating soft or semisoft materials.
WD-08517-65
J
–90 to 760°C / –310 to 1400°F
25 seconds (in liquids)
WD-08516-65
K
–250 to 899°C / –418 to 1500°F
25 seconds (in liquids)
WD-08500-65
T
–250 to 400°C / –418 to 752°F
25 seconds (in liquids)
Air/gas probes.
Feature a perforated shield to minimize error from radiated heat.
WD-08517-75
J
–90 to 760°C / –310 to 1400°F
225 seconds at 5 m/s airflow
WD-08516-75
K
–250 to 899°C / –418 to 1500°F
225 seconds at 5 m/s airflow
WD-08500-75
T
–250 to 400°C / –418 to 752°F
225 seconds at 5 m/s airflow
Surface probes.
Use for surfaces as hot plates, furnaces and molds. Exposed junction is isolated from 316 SS shaft
and aluminum housing with ceramic support.
WD-08517-60
J
–90 to 760°C / –310 to 1400°F
30 seconds
WD-08516-60
K
–250 to 899°C / –418 to 1500°F
30 seconds
WD-08500-60
T
–250 to 400°C / –418 to 752°F
30 seconds
Thermocouple Probes
All probes listed below include an integral handle and a 5-ft coiled cable with a miniconnector.
General purpose and penetration probes feature a grounded junction for fast response. Shpg wt 0.2 lb (0.1 kg)
See bulletin TEMP B1 for information on OAKTON ACORN thermistor and RTD thermometers.
NIST-Traceable Calibration and Certificates are available;
please contact your authorized OAKTON distibutor for more information.