- 6 - 

Table-2

 

Calibration icon

 

Conductivity Standards

 

 

Oakton part numbers 

 

500 mL bottles, 20 mL pouches(20/box)

 

4-Point 

calibration

 

 

84 μS/cm

 

 

00653-16, 35653-08 (Pouches)

 

 

1413 μS/cm

 

 

00653-18, 35653-11 (Pouches)

 

 

12.88 mS/cm

 

 

00606-10, 35653-14 (Pouches)

 

111.8 mS/cm

 

 

 

4.3.2  How often to calibrate 

 

We recommend calibrating the conductivity meter once per month under normal conditions.

 

 

If the requirement for accuracy is high or the temperature of the test sample has a big difference to 

the reference temperature (25 ˚C), we recommend doing the calibration at least once a week.

 

 

Users can tell if the meter needs a calibration by testing the standard calibration solutions. If the 

measurements you get have a big difference to the standard value of the calibration solutions, then a 

calibration is necessary.

 

 

For new electrodes or the meter that has been restored to factory default setting, we recommend 

conducting 3-4 points calibration for best accuracy. For general uses, users can choose the 

calibration solutions that are close to the test samples to conduct 1 or 2 points calibrations.

 

 

4.4 

Conductivity Calibration 

(take 1413 

μ

S as an 

example)

 

4.4.1.

 Rinse the electrode in distilled or pure water, allow it to dry, wash with 

a  little  of  standard  solution  and  submerge  it  in  standard  solution.  Stir  the 

solution briefly and allow it to stay in the solution until a stable reading is 

reached. 

 

4.4.2.

 

Press 

  key  to  enter  the  calibration  mode. The  “

CAL

”  icon  will 

flash  at the top right, and the  scanning and locking process of calibration 

solution will appear at the bottom right. 

 

4.4.3.

 When the meter locks 1413 μS, stable icon 

  will display on LCD. 

Press 

  key to complete calibration. 

“End”

 icon will show up and the 

meter will return to measuring mode and 

  will show up on the bottom 

left (see Figure-7).

 

Note: 

pressing 

  before the reading being stable (

shows up and stays), the 

Err4

 self-diagnosis icon will appear (see section 6.5.2).

 

4.4.4.

 For multi-point calibration, please repeat clause 4.3.1-4.3.2 until all the calibration is done. The 

meter can repeat calibration in the same calibration solution until the stable value is reached.

 

Figure-7

 

Summary of Contents for CON550

Page 1: ...CON550 Benchtop Conductivity Meter Instruction Manual OAKTON INSTRUMENTS www 4oakton com v2 0...

Page 2: ...e 4 3 5 Data storage recall and deletion 4 3 6 Max Min Function 5 4 Conductivity Measurement 5 4 1 Preparation 5 4 2 Conductivity Electrode Information 5 4 3 Information regarding Calibration 5 4 4 Co...

Page 3: ...the meter s response time and measuring accuracy with stable reading display mode available The kit comes with a combination conductivity electrode with temperature sensor measuring conductivity and t...

Page 4: ...dustproof and spillproof Size and Dimension 240 235 103 mm 1kg 3 Instrument Description 3 1 LCD Display 1 Measurement mode 2 Measurement value 3 Calibration icons and numberings 4 Measurement unit 5 T...

Page 5: ...suring mode Cond TDS Long Press Enter parameter setting P1 P2 P6 Short Press Enter Calibration mode Cancel any operation and go back to measurement mode Short Press In calibration mode press to conduc...

Page 6: ...st group of the measuring data Press again to keep storing 3 5 2 Recall stored data In the measurement mode long press key to recall the last stored Measuring data The meter displays RM icon and stora...

Page 7: ...pter Plug the power adapter tightly into the DC9V socket The power s voltage should meet the requirements listed on the power adapter 4 2 Conductivity Electrode Information 4 2 1 Conductivity Electrod...

Page 8: ...users can choose the calibration solutions that are close to the test samples to conduct 1 or 2 points calibrations 4 4 Conductivity Calibration take 1413 S as an example 4 4 1 Rinse the electrode in...

Page 9: ...of the solution 4 6 2 During the process of calibration and measurement the meter has self diagnosis functions indicating the relative information as below Table 8 Table 8 Self diagnosis information o...

Page 10: ...ting P1 Then press or to change menu from P1 to P2 P7 For details please see Table 10 Table 10 pH Parameter Setting P1 Select electrode s constant 1 0 10 0 0 1 1 Long press to enter P1 2 Press 1 0 fla...

Page 11: ...lean Before taking a measurement rinse the electrode in pure water It is recommended to rinse it again in the sample solution When submerge the electrode in solution stir the solution briefly to elimi...

Page 12: ...ENTS any malfunctioned or damaged product attributable to responsibility of OAKTON INSTRUMENTS for a period of three years from the delivery a six month limited warranty applies to probes This warrant...

Reviews: