background image

 

- 7 - 

4.5 

TDS and Conductivity

 

4.5.1.

  TDS  and  conductivity  are  linear  related.  The  conversion  factor  is  0.40-1.00. Adjust  the  factor  from 

parameter P2.6. The factory default setting is 0.71 and please refer to Section 7.4. The meter only needs to 

be calibrated in Conductivity mode, then after calibration of conductivity, the meter can switch from conductivity 

to TDS.

 

4.5.2.

 Adjust TDS conversion factor in parameter setting P4 according to the data collected during testing and 

experience. Table – 10 lists some commonly used Conductivity and TDS conversion factors. This is for your 

reference only.

 

Table – 10 Conductivity and TDS conversion factors 

 

Conductivity of solution

 

TDS conversion factor

 

0-100 μS/cm

 

0.60

 

100-1,000 μS/cm

 

0.71

 

1-10 mS/cm

 

0.81

 

10-100 mS/cm

 

0.94

 

 

4.6 

Sample Test

 

4.6.1.

 Rinse conductivity electrode in distilled or pure water, dry it, and submerge it in the sample solution. Stir 

the solution briefly and allow it to stay in the sample solution until a stable reading is reached and 

 

icon 

appears on LCD, then get the reading value, which is the conductivity value of the solution.

 

4.6.2.

 During the process of calibration and measurement, the meter has self-diagnosis functions, indicating 

the relative information as below: Table – 8.

 

Table – 8 Self-diagnosis information of conductivity measurement mode

 

Display Icons

 

Contents

 

Checking

 

 

Wrong conductivity calibration solution or 
the meter recognition of calibration 
solution is out of range.

 

1. Check whether conductivity calibration 

 

solution is correct.

 

2. Check whether the meter connects the 

 

electrode well.

 

3. Check whether the electrode is damaged.

 

 

 

Press 

 

key when measuring value is 

not stable during calibration.

 

 

Press 

 

key

 

after 

 

icon appears.

 

 

During  calibration,  the  measuring  value 
being unstable for over 3 minutes. 

 

1.  Shake  the  electrode  to  eliminate  bubbles  in 

electrode head.

 

2. Replace with new conductivity electrode.

 

 

4.6.3.

 Factory default setting

 

For factory default setting, please refer to parameter setting P7 (Section 4.7.2). With this function, all calibration 

data is deleted, and the meter restores to the theory value. Some functions restore to the original value. When 

calibration  or  measurement  fails,  please  restore  the  meter  to  factory  default  setting  and  then  perform  re-

calibration or measurement. Please note once set the factory default, all the data deleted will be irretrievable. 

 

Summary of Contents for CON550

Page 1: ...CON550 Benchtop Conductivity Meter Instruction Manual OAKTON INSTRUMENTS www 4oakton com v2 0...

Page 2: ...e 4 3 5 Data storage recall and deletion 4 3 6 Max Min Function 5 4 Conductivity Measurement 5 4 1 Preparation 5 4 2 Conductivity Electrode Information 5 4 3 Information regarding Calibration 5 4 4 Co...

Page 3: ...the meter s response time and measuring accuracy with stable reading display mode available The kit comes with a combination conductivity electrode with temperature sensor measuring conductivity and t...

Page 4: ...dustproof and spillproof Size and Dimension 240 235 103 mm 1kg 3 Instrument Description 3 1 LCD Display 1 Measurement mode 2 Measurement value 3 Calibration icons and numberings 4 Measurement unit 5 T...

Page 5: ...suring mode Cond TDS Long Press Enter parameter setting P1 P2 P6 Short Press Enter Calibration mode Cancel any operation and go back to measurement mode Short Press In calibration mode press to conduc...

Page 6: ...st group of the measuring data Press again to keep storing 3 5 2 Recall stored data In the measurement mode long press key to recall the last stored Measuring data The meter displays RM icon and stora...

Page 7: ...pter Plug the power adapter tightly into the DC9V socket The power s voltage should meet the requirements listed on the power adapter 4 2 Conductivity Electrode Information 4 2 1 Conductivity Electrod...

Page 8: ...users can choose the calibration solutions that are close to the test samples to conduct 1 or 2 points calibrations 4 4 Conductivity Calibration take 1413 S as an example 4 4 1 Rinse the electrode in...

Page 9: ...of the solution 4 6 2 During the process of calibration and measurement the meter has self diagnosis functions indicating the relative information as below Table 8 Table 8 Self diagnosis information o...

Page 10: ...ting P1 Then press or to change menu from P1 to P2 P7 For details please see Table 10 Table 10 pH Parameter Setting P1 Select electrode s constant 1 0 10 0 0 1 1 Long press to enter P1 2 Press 1 0 fla...

Page 11: ...lean Before taking a measurement rinse the electrode in pure water It is recommended to rinse it again in the sample solution When submerge the electrode in solution stir the solution briefly to elimi...

Page 12: ...ENTS any malfunctioned or damaged product attributable to responsibility of OAKTON INSTRUMENTS for a period of three years from the delivery a six month limited warranty applies to probes This warrant...

Reviews: