Instruction Manual PC 300
21
Taking measurements with READY indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the reading is
stable
*. Switch the READY indicator on or off in SETUP program P3.1. See Section 7.4.
* The READY indicator appears and the reading holds until the measured value exceeds the
tolerance (
±
0.02 pH;
±
0.8 mV <400;
±
1.2 mV > 400). Then, the READY annunciator turns
off.
Taking measurements with the auto endpoint feature selected on
When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically
“HOLD” the reading. The “HOLD” indicator appears on the left side of the display. Press the
HOLD
key to release the reading. Switch the Auto endpoint feature on or off in SETUP
program P 3.1, s
ee Section 7.4.