Instruction Manual
PC 300
61
B. Press
the key until the lower display reads P8.0.
C. Press
the
ENTER
key twice. The lower display reads tCO and the upper display
shows the temperature coefficient value (should be 0, as per step 1 above).
D. Press
the
key until the upper display shows your calculated temperature
coefficient.
E. Press
ENTER
key to confirm the value.
F. Press
CAL/MEAS
key twice to return to measurement mode.
The calculated temperature coefficient will not be applied to all the meter readings.