Instruction Manual
PD 300
P3.2: Selecting number of pH calibration points
Program P3.2 lets you select the number of calibration
points that appear in pH calibration mode: 2, 3, 4, or 5.
The meter will automatically exit calibration mode after
you have calibrated to your selected number of points.
From measurement mode.
1. Press
the
MODE
key to select pH measurement
mode.
2.
Press SETUP key to enter Set Up mode.
3. Press
the
▲
or
▼
keys to scroll through subgroups
until you view parameter P3.0.
4.
Press the ENTER key twice to select parameter 3.2.
5. Press
the
▲
or
▼
keys to select 2, 3, 4, or 5 point
pH calibration.
6. Press the ENTER key to confirm selection and to
return to the subgroup menu. Press CAL/MEAS
key to return to measurement mode.
See Figure on right.
Figure 25 – P3.2: Select number
of pH calibration points
34