Instruction Manual
PD 300
8.12
P7.0 [6.0] rSt: Resetting to Factory Default Setting (DO)
This program lets you reset all DO (% saturation and mg/L) parameters to factory default
settings. This clears all calibration data and any other DO setup functions you might have
changed.
The following settings will remain as you have set them:
•
Temperature unit of measure (°C or °F)
•
The temperature offset calibration value
•
All DO calibration data and parameters
There are also some other parameters that retain settings when reset is done. See Addendum
3 on page 68 for a table of factory default settings.
From measurement mode
Figure 41: Reset to factory
defaults
1.
Press SETUP key to enter SETUP mode.
2. Press
▲
or
▼
keys to scroll through subgroups
until you view parameter “rSt” in the upper display.
3.
Press the ENTER key.
4. Press
▲
or
▼
keys to toggle between NO and
YES.
•
NO retains current memory.
•
YES clears all memory
5. Press the ENTER key to confirm selection and to
return to measurement mode.
49