16. CALIBRATION SCREEN
The thermometer is factory calibrated and
does not require calibration before use.
The Calibration function allows single
point calibration of the thermometer, at
0°C (32°F) to compensate for
thermocouple off-set error. It is NOT
necessary to perform a field calibration to
obtain specified meter accuracy. Use the
field calibration feature to improve
thermometer/probe accuracy or to
compensate for thermocouple drift...
Before go into the calibration mode, must
enter the password. Press F2 to change to
the next digit. (Default Password is 9900)
There are three calibration options:
Offset – adjusts at a single point
Slope – adjusts at two points
Match – adjusts readings on T1 to match
those on T2. Or adjusts T2 to match T1.