Chapter 2
OaktonLog Plus
152
2.5.5.
Sensor Calibration
A new OaktonLog comes fully calibrated. After a period of long
use, however, you may want to re-calibrate the humidity or the
temperature sensors.
The humidity calibration affects both OaktonLog and OaktonLog
plus and should be carried out while OaktonLog is connected to the
PC. The temperature calibration affects only the OaktonLog Plus
software and can be carried out regardless of whether OaktonLog
is connected to the PC.
1.
Calibration password
To prevent accidental change of the calibration, the calibration
procedure is protected by a password. The default password is:
1234. To change the password: