Chapter 2
OaktonLog Plus
156
7.
Disconnect OaktonLog from the PC and insert it into the
humidity chamber.
8.
Set the humidity chamber to the first reference value.
Wait until the humidity level is stabilized and write down
the OaktonLog’s reading.
9.
Repeat the last step with the second reference value.
10.
Connect the OaktonLog to the PC.
11.
Enter the two OaktonLog values into the OaktonLog
Value edit boxes.
12.
Enter the two reference values into the Reference Value
edit boxes.
13.
Click
Calibrate
.
14.
To restore default calibration, click
Default
.