- 5 -
Ingress protection:
Meets IEC-529 IP-54 for dust and water resistant enclosures (probe attached)
CE Compliance
EN61326-1/A1: 1998 (EU EMC Directive)
Page 1: ...al Temp 360 RTD Datalogging Thermometer 68X526601 Rev 1 09 09 Part of Thermo Fisher Scientific 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEqu...
Page 2: ......
Page 3: ...DISPLAY OVERVIEW 11 10 MEASUREMENT MODE 13 11 HOLD FUNCTON 14 12 MIN MAX and AVE FUNCTION 14 13 DATA LOGGING 14 14 SETUP MODE 15 15 GENERAL SETUP SCREEN 16 16 CALIBRATION SCREEN 18 17 ALARMS SCREEN 20...
Page 4: ......
Page 5: ...atalogging for up to 2000 points USB output Operator selection of Celsius or Fahrenheit scale Resolution of up to 0 01 C F Large backlit LCD and dot matrix graphic LCD display Hold feature for tempora...
Page 6: ...zardous atmosphere batteries must only be changed in an area known to be non hazardous CAUTION 1 Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas 2 Weak batte...
Page 7: ...01 C F From 99 99 to 99 99 C F 0 1 C F From 100 0 to 999 9 C F 1 C F Above 1000 C F Accuracy 0 1 C 0 2 F From 330 0 to 100 C F 0 03 C 0 06 F From 99 99 to 99 99 C F 0 1 C 0 2 F From 100 0 to 999 9 C F...
Page 8: ...pin DIN connector Input Protection 50V rms Storage 40 C to 65 C 40 F to 149 F Humidity 10 to 90 non condensing Battery Life Size Three AA 1 5V Alkaline Life 400 hours continuous typical without backli...
Page 9: ...5 Ingress protection Meets IEC 529 IP 54 for dust and water resistant enclosures probe attached CE Compliance EN61326 1 A1 1998 EU EMC Directive...
Page 10: ...mpty bars flashing More than 2 85 V Adaptor power supply Main power supply connected USB power supply USB power Selected settings are stored in memory and will remain in memory even after power is tur...
Page 11: ...r armor slide in from the top of meter before pushing the bottom edges of meter down to set it into position Lift up the stand at the back of meter for bench top applications if necessary 2 To remove...
Page 12: ...8 6 ASSEMBLING OPTIONAL HANDSFREE ACCESORIES You can use the optional magnets and strap in the Handsfree Kit accessories for hands free operations...
Page 13: ...ings Insert the 3 pin plug into the mating connector on the top of the instrument If no probe is connected the display will read open RTDs are sensitive at the tip or sensing element When taking measu...
Page 14: ...Step through Min Max and Avg readings F2 Toggle between F and C display F3 Toggle between menu and measure mode hold Freeze display on off light Turns meter on and off press and hold for 3 seconds to...
Page 15: ...11 9 DISPLAY OVERVIEW The dot matrix display features a large primary display smaller secondary displays for channel info or min max ave and helpful annuciators for added measurement data...
Page 16: ...icator 8 Data logging indicator 9 PC data logging indicator 10 Main reading display 11 Current reading unit indicator 12 Min Max Avg elapsed time 13 Current Min Max Avg reading indicator 14 Meter logg...
Page 17: ...will display the measured value for in the primary display Pressing the F2 key will toggle reading between F and C display Pressing the F1 key initiates and toggles through Minimum Maximum and Averag...
Page 18: ...re change that occurs The minimum and maximum values are sensed and automatically stored To exit and clear this function press the F3 to access the Menu functions See the Clear Reset menu section for...
Page 19: ...e meter key pad to scroll through options To enter a setup screen press Select F1 key To return to measurement mode press Meas F3 key Following menu options are listed 1 General Setup 2 User field cal...
Page 20: ...ion auto off time and password Press F1 to indicate you want to change the setting of the current parameter or recall or log to move to the next parameter Press recall or log to change the options Pre...
Page 21: ...17 This screen below is used to reset change password In the event if uses forget his her password 5586 can be used to reset to a new value On the second page you can set time and date...
Page 22: ...meter at any temperature point to compensate for RTD off set error It is NOT necessary to perform a field calibration to obtain specified meter accuracy Use the field calibration feature to improve th...
Page 23: ...rature in the offset range of 10 00 C or 18 00 F Slope adjusts at two points The two calibration points must be at least 40 00 C 72 00 F apart The 2nd calibration point should be at a higher value tha...
Page 24: ...nd of alarm setting is available under the alarm setting options Measurement alarm Disable or enable the alarm by pressing recall or log and F1 to accept Increase or decrease individual limit by press...
Page 25: ...21 Countdown alarm You can enable disable the countdown alarm and set the countdown time from 5 sec to 1 hour After setting enabling the countdown alarm the measurement screen should look like this...
Page 26: ...ountdown Press F2 key in the middle of a countdown will stop the process You will need to restart the countdown by selecting the start key F2 Alternatively you can also choose Menu to go into the time...
Page 27: ...the timer select restart F2 it You display should look like this To repeat the functions select F2 Note The C F function is disabled when the count down function is enabled To activate the o C o F fu...
Page 28: ...e countdown timer temporarily over rides the Auto shutoff until the countdown is completed or manually stopped If the meter is manually or auto shutoff The Countdown Alarm is automatically reset to Di...
Page 29: ...as auto or manual If it is auto logging using recall or log to set time interval Its range is from 2min to 60min Data Transfer from Meter to Computer Once the USB connection is establish with PC pres...
Page 30: ...26 18 CALIBRATION REPORT SCREEN The Calibration report will show the time and date along with results of the last user calibration...
Page 31: ...27 19 CLEAR RESET SCREEN Press F1 to choose which data you want to clear or reset For calibration logged data and reset all you will have to enter the password to proceed Default Password is 9900...
Page 32: ...BY ELECTROSTATIC DISCHARGE CLEAN WITH DAMP CLOTH Do not clean with abrasives or solvents Use mild detergents never immerse nor use excessive fluid 22 BATTERIES If there is no display when the thermome...
Page 33: ...FAULT ACTION NO Display Check condition of batteries Check that batteries are inserted properly Display shows OVER UNDER Out of range indication Display Shows OPEN No RTD connected in the Connector D...
Page 34: ...ermometer 35426 40 Rubber Armour with Stand 35427 80 Handsfree Kit Two Magnets and a Strap 35427 85 General purpose probe immersion Into liquids 08117 70 Penetration probe meat semi soft Materials 081...
Page 35: ...s outside the warranty period or those resulting from product misuse or abuse may be invoiced to you 26 PRODUCT RETURN To limit charges and delays contact the seller or Manufacturer for authorization...
Page 36: ...tomers meet ISO FDA EPA GLPs cGMPs and other quality standards Conformity ISO IEC 17025 2005 accredited NIST Handbook 150 2000 Edition ANSI NCSL Z540 2 1997 NIST Technical Note 1297 ISO 9000 2000 Fast...
Page 37: ...ce your instrument We provide you with the documentation you need to meet your most stringent quality requirements for the control of inspection measuring and test equipment Certification includes cer...
Page 38: ...send you an automatic reminder when it is time to recalibrate your instrument Are your calibration certificates good enough InnoCal surpasses the competition by providing the most complete certificat...
Page 39: ...ceable standards Meter only Probe only System meter probe Four test points across range of instrument 0 100 165 230 C 4 32 446 770F MM 17000 04 MM 17001 04 MM 17002 04 InnoCal The Choice of Quality 86...
Page 40: ...36...
Page 41: ......
Page 42: ...facturer or authorized seller For more information on OAKTON Instruments Products please contact your nearest distributor or visit our web site listed below Distributed by 99 Washington Street Melrose...