background image

17

CLEANING & DISINFECTION / INSPECTIONS / WARRANTY

CLEANING PROCESS

To reduce the risk of electric shock:

-  Always unplug this furnishing form the electrical outlet before cleaning.

DANGER

Follow the cleaners/disinfectant manufacturers’ directions for use. Please note that cleaning and disinfecting an OAKWORKS

®

 table 

is a two part process. First it must be cleaned of any visible soil, then it can be disinfected. The table should be cleaned as soon as 

possible after use. Please follow these procedures for best results:

1.  Using a fresh wipe or wipes, clean any visible soil off of the table, working from the top to the bottom of the table.
2.  Throw the used wipe(s) away.
3.  Using a fresh wipe or wipes, thoroughly wipe all surfaces of the tabletop and any high-contact areas such as handles, hand-

sets, etc., making sure they remain wet for the disinfectant manufacturer’s recommended contact time.

4.  Throw the used wipe(s) away and allow the table to dry.

INSPECTIONS

RECOMMENDED REGULAR INSPECTIONS (monthly or local standard)

• 

Check for damage to the power, hand control(s) and foot control(s) on all cables.

• 

Visually inspect components for obvious damage that could cause problems during operation.

RECOMMENDED PERIODIC INSPECTIONS (yearly or local standard)

• 

Check for damage to the power, hands control and foot control cables and all visible wiring.

• Visually inspect components for obvious damage that could cause problems during operation.
• Check all mechanical functions using the hand control. Repeat using the foot control. Check for abnormal noises.
• Check that all fasteners are present and fastened securely.
• Check table grounding.
• Clean unusual buildup of dirt on the table and/or parts of the table not normally cleaned on a regular basis.
• Check for tears or cracks in the upholstery.

WARRANTY

View complete warranty details at www.oakworks.com

Summary of Contents for PF Series

Page 1: ...L www oakworksmed com 717 235 6807 made in the USA with US imported parts PF250 PF200 Shown with optional QuickLock Face rest platform and Boiance crescent PF400 Shown with optional Adjustable Side Ar...

Page 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifications and the like documen...

Page 3: ...rest Adjustment 12 Lowered Arm Rest Use 12 Locking Casters Use 13 Table Extender 13 Adjustable Arm Rest Installation and Adjustment 14 QuickLock Face Rest 15 Cleaning Disinfection Recommended Cleaners...

Page 4: ...Terminal This symbol when used in this manual or on product labels indicates that the product should be protected from moisture The humidity specifications for Transport Storage are listed in the env...

Page 5: ...thetic mixture with air or oxygen or nitrous oxide Do not lift the table by the table top This can damage the table When lowering the table or using the Trendelenburg functions make sure there is noth...

Page 6: ...0 kg patient and accessories Widths 27 69 cm 29 74 cm 31 79 cm or 33 84 cm Length 4 section 72 183 cm Padding 4 10 cm Comfort Foam Upholstery Terra Touch Wood Stain Natural Maple Finish Table Weight 2...

Page 7: ...s 250 kg patient and accessories Widths 27 69 cm 29 74 cm 31 79 cm or 33 84 cm Length 2 section 74 188 cm Padding 3 8 cm Comfort Foam Upholstery Terra Touch Wood Stain Natural Maple Finish Table Weigh...

Page 8: ...patient and accessories Widths 27 69 cm 29 74 cm 31 79 cm or 33 84 cm Length 2 section 73 185 cm Padding 3 8 cm Comfort Foam Upholstery Terra Touch Wood Stain Natural Maple Finish Table Weight 212 lb...

Page 9: ...is equipped with a cord having an equipment grounding conductor and a grounding pin The pin must be plugged into an appropriate outlet that is prop erly installed and grounded in accordance with all l...

Page 10: ...anything Damage can occur to the table The PF400 Tables offer trendelenburg height backrest and footrest positioning with electronic ease Operate the foot control as shown below to raise or lower the...

Page 11: ...or holding anything Damage can occur to the table The PF250 PF200 tables offer trendelenburg and height positioning with electronic ease Operate the foot control as shown below to raise or lower the t...

Page 12: ...his area for holding anything Damage can occur to the table The PF400 Tables offer trendelenburg height backrest and footrest positioning with electronic ease Operate the optional hand control as show...

Page 13: ...nder the backrest top is not designed for storage DO NOT use this area for holding anything Damage can occur to the table The PF250 PF200 tables offer trendelenburg and height positioning with electro...

Page 14: ...use the table to transport a patient 11 DIRECTIONS FOR USE Grasp at base corner at low height range 1 2 13mm Open end Wrench Leveler Foot 1 Place the table in the location where you plan to use it 2...

Page 15: ...side of the back rest by removing the screws and relocating the hand lever to the opposite side of the top into the four pre drilled holes Back Rest Control The Lowered Arm Rest can be removed when no...

Page 16: ...CKING THE CASTERS 1 Place your shoe on top of the casters locking tab 2 Press firmly inward and downward until you here a click ing noise and the caster is unlocked CAUTION TABLE EXTENDER USE Do not p...

Page 17: ...rm rest at a 60 angle to the backrest 2 Push end all the way into the hole and drop the arm rest into the horizontal position Pull outward to make sure the arm rest is secure TO LEVEL THE ARM RESTS Th...

Page 18: ...ting for height and angle and then locking the cam locks into the desired position Important Safety Note Maximum Distance from edge of Table 3 7 6 cm QuickLock Face Rest Insert dowels into Accessory a...

Page 19: ...cquer plated metals 2 Clorox Hydrogen Peroxide Disinfectant EPA reg 67619 24 corrosion of metals deterioration of upholstery and wood lacquer leaves a hard to remove residue 3 Virox Accel TB EPA reg 7...

Page 20: ...ain wet for the disinfectant manufacturer s recommended contact time 4 Throw the used wipe s away and allow the table to dry INSPECTIONS RECOMMENDED REGULAR INSPECTIONS monthly or local standard Check...

Page 21: ...Series Exam Table Exam Base Top FT Flat PF200 LB Lift Assist PF250 ES Electric PF400 Table Width Minimum Table Height Table Top Length Padding PL 3 AC 4 Fabric TerraTouch Grommets Options Model Number...

Page 22: ...ICAL SPECIFICATIONS Designed for North America Europe Input Service 120 VAC 15 amp 60 Hz 220 VAC 10 amp 50 60 Hz Current Draw 4 0 amps 2 0 amps Maximum Momentary Current Consumption 6 0 amps 3 0 amps...

Page 23: ...netic environment in which radiated RF disturbances are controlled the customer or the user of the table can help prevent electromagnetic interference by maintaining a minimum distance between portabl...

Page 24: ...de 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical com mercial or hospital environ ment Voltage dips short inter ruptions and voltage variation...

Page 25: ...RF transmitters as deter mined by an electromagnetic site survey ashould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with th...

Page 26: ...23 THIS PAGE IS INTENTIONALLY LEFT BLANK...

Page 27: ...24 THIS PAGE IS INTENTIONALLY LEFT BLANK...

Page 28: ...m PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com made in the USA with US imported parts 3034177 CONFORMS TO ANSI AAMI STD ES60601 1 IEC STD 60601 1 3RD EDITION IEC STD 60601 1 2 3RD...