background image

USER MANUAL

www.oakworksmed.com  ·  717.235.6807

made in the USA

with US & imported parts

OAKWORKS

®

 

DTPM 300 Imaging Table

Summary of Contents for DTPM 300

Page 1: ...USER MANUAL www oakworksmed com 717 235 6807 made in the USA with US imported parts OAKWORKS DTPM 300 Imaging Table...

Page 2: ...within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS Inc OAKWORKS Inc encourages requests for technical specifications and the like documen...

Page 3: ...Fiber Arm Board 10 Fluoro Extender 10 Spine Positioning System II 11 Locking Casters Use 11 Reference Usable Imaging Area Diagrams 12 13 Cleaning Disinfection Recommended Cleaners Disinfectants 14 Cle...

Page 4: ...is not designed for and should not be used with Magnetic Resonance Imaging procedures Although the table accommodates a number of procedures it is not intended to serve as a surgical procedure table S...

Page 5: ...ed Both locking knobs must be securely tightened at all times Travel can occur if the locking knobs are not secured by inadvertently pushing on the end of the table or by tilting the table The table u...

Page 6: ...ent exposure to scattered radiation from the x ray generat ing source Oakworks Imaging Tables may be used for x ray imaging where the x ray generator is located below the tabletop and the image recept...

Page 7: ...ge Op tions 120V 60Hz or 230V 50Hz Table Capacity 450 lbs 205 kg patient and accessories Polycarbonate Top 78 198 cm Long x 24 61 cm Wide 4 10 cm Locking Casters Locking Unlocking Casters for mobile o...

Page 8: ...uct is equipped with a cord having an equipment grounding conductor and a grounding pin The pin must be plugged into an appropriate outlet that is proper ly installed and grounded in accordance with a...

Page 9: ...peutic procedures DANGER Should the use of this device create a circumstance under which the patient could be overexposed to ionizing radiation discontinue use immediately and determine an alternative...

Page 10: ...all cautions and warnings Reverse Trendelenburg activates tilt up to 14 positioning stops when table is level press again for complete reverse Trendelenburg activates tilt up to 14 positioning stops w...

Page 11: ...ide of the table just under the side rail Rotate the locking knob several turns counter clockwise to loosen Slide the table in the desired direction Rotate the locking knob clockwise until tight to se...

Page 12: ...hen pressing the flaps in place anchoring the flaps to the Fastener Hook and Loop Faster as shown applies to both the 1 2 5 cm 2 5 cm pads Hook and Loop Faster All electric tables are equipped with an...

Page 13: ...cedure The weight of the patient will hold the Arm Board in place Rotate the Support Section to the desired position This section rotates up to 180 Prone Position at 135 Supine Position at 180 Support...

Page 14: ...ARGE RECTANGULAR ADJUSTER PAD 7 X 12 X 1 18cm x 31cm x 4cm SMALL RECTANGULAR ADJUSTER PAD 8 20 CM SEMI ROUND BOLSTER All four casters should be locked at all times during use with a patient LOCKING CA...

Page 15: ...M 37 93 96 33 CM 16 57 CM 12 01 30 51 CM 12 01 30 52 CM 6 52 16 57 CM 37 93 96 33 CM 6 52 16 57 CM 12 01 30 51 CM 16 34 41 50 CM 19 88 50 48 CM 7 85 19 93 CM 7 02 17 84 CM 34 49 87 61 CM 7 02 17 84 CM...

Page 16: ...13 REFERENCE OFFSET FOR A P VIEW OFFSET CAUDAL 30 VIEW FROM VERTICAL NO IMAGE NO IMAGE NO IMAGE NO IMAGE...

Page 17: ...cquer plated metals 2 Clorox Hydrogen Peroxide Disinfectant EPA reg 67619 24 corrosion of metals deterioration of upholstery and wood lacquer leaves a hard to remove residue 3 Virox Accel TB EPA reg 7...

Page 18: ...ain wet for the disinfectant manufacturer s recommended contact time 4 Throw the used wipe s away and allow the table to dry INSPECTIONS RECOMMENDED REGULAR INSPECTIONS monthly or local standard Check...

Page 19: ...CRIPTION FL DT RT 22 26 84 MG TT XX XX Fluoroscopy Table Base Motion Dual Tower Top Polycarbonate RT Rectangular Table Top Width Minimum Table Height Table Top Length Padding Comfort Foam Fabric Terra...

Page 20: ...e standards 21 CFR Subchapter I in effect at the time of manufacture for radiographic tables ENVIRONMENTAL CONDITIONS Conditions Temperature Humidity Atmospheric Pressure Normal Use 50 10 C to 104 40...

Page 21: ...magnetic environment in which radiated RF disturbances are con trolled the customer or the user of the table can help prevent electromagnetic interference by maintaining a minimum distance between por...

Page 22: ...2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrup tions and voltage variati...

Page 23: ...d RF transmitters as deter mined by an electromagnetic site survey ashould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with t...

Page 24: ...g Road New Freedom PA 17349 Phone 717 235 6807 FAX 717 235 6798 www oakworksmed com made in the USA with US imported parts 3034177 CONFORMS TO ANSI AAMI STD ES60601 1 IEC STD 60601 1 3RD EDITION IEC S...

Reviews: