17
SPECIFICATIONS
PRODUCT SPECIFICATIONS
Materials of Construction
This product contains no latex.
Top Frame
Extruded aluminum with powdercoated or anodized surfaces.
Table Top
.5” (1.3 cm.) Polycarbonate
Patient Comfort
1” (2.5 cm.) upholstered foam pad
Table Length
78” (198 cm.)
Table Width
24” (61 cm.)
Height Range
24”-40” (61-102 cm.) standard
Table Top Travel Range
±11.5” (29 cm)
Trendelenburg Tilt
±14°
Weight
390 lbs. (177 kg.)
Shipping Weight
526 lbs. (239 kg.)
Lifting Capacity
450 lbs. (205 kg.)
Aluminum Filtration Equiva-
lent of Table top
TABLE TOP
100 kvp, Hvl of 3.6 mm = 1.62 (1.74 max)
1” (2.5 cm.) TABLE TOP PAD
100 kvp, Hvl of 3.6 mm = .88 (1.07 max)
2” (5 cm.) TABLE TOP PAD
100 kvp, Hvl of 3.6 mm = 1.47 (1.76 max)
This product complies with the United States Department of Health and Human Services radiation performance
standards, 21 CFR Subchapter I, in effect at the time of manufacture for radiographic tables.
ENVIRONMENTAL CONDITIONS
Conditions
Temperature
Humidity
Atmospheric Pressure
Normal Use
50° (10°C) to 104° (40°C)
20% to 60% RH
98 to 105 kPa
Storage & Transport
-20° (-29°C) to 135° (57°C)
20% to 95% RH
98 to 105 kPa
ELECTRICAL SPECIFICATIONS
Designed for:
North America
Europe
Japan
Input Service
120 VAC/15 amp/60 Hz
220 VAC/10 amp/50/60 Hz
100 VAC/20 amp/50 Hz
Current Draw
5.8 amps
2.9 amps
4.8 amps
Maximum Momentary Current
Consumption
6.0 amps
3.0 amps
7.2 amps
Voltage Output to Actuators
24 VDC
24 VDC
24 VDC
Electric Shock Protection
Class 1 Equipment
Class 1 Equipment
Class 1 Equipment
Tabletop Applied Part
Type B Applied Part
Type B Applied Part
Type B Applied Part
Ingress Protection Rating
IPX0
IPX0
IPX0
Made of Operation
Intermittent Operation
MAX 2 minutes ON
18 minutes off
Intermittent Operation
MAX 2 minutes ON
18 minutes off
Intermittent Operation
MAX 2 minutes ON
18 minutes off