25
TROUBLESHOOTING
TROUBLESHOOTING
TROUBLESHOOTING
• Is the unit plugged in?
• Does the outlet have power?
• Check all connections of the control box, is everything seated securely?
• Does the height of the table work when you press on the foot pedal?
• Does the top work when you use the hand control?
IF THE TABLE WILL NOT CHANGE POSITION
CAUTION
The table weighs as much
as 300 lbs. DO NOT lift or
move the unit without using
proper safety procedures
or equipment. DO NOT turn
the table
on its side for any
reason. To do so could cause
damage to the base and
will void the warranty. There
are no serviceable items that
need this action.
IMPORTANT
Be sure to read all
cautions, warnings
and instructions given
in the manual to
prevent injury to both
operator and client.
•
Check the outlet to be sure that it has power and that the power cable is plugged
in. Try each foot control and hand control to see if any work. If any work, then
possibly there is a loose connection or a faulty controller.
•
It may be necessary to do a reset of the control box.
a. Press second row of 2 buttons simultaneously on handset
(simetimes it takes several attempts to press them “simultaneously”)
b. Control box will beep approximately 10 times
c. Run all motors in system to fully retracted position (down)
d. Re-test table to insure proper functions.
•
Unplug the power cable from the wall outlet.
•
Make sure that all male and female connectors on the control box
are firmly pushed into each other. Even a slight misconnection can
cause a possible malfunction. Check the connections on top at the
columns.
•
Plug the unit back into the wall and check the controls again. If none work, then
contact customer service.
QUESTIONS TO ASK BEFORE CALLING CUSTOMER SERVICE
The Oakworks Ultrasound tables have two lifting columns that raise and lower the table.
In addition, the Multi-Specialty & Women’s Imaging have one actuator that allow the
client to be positioned for both client and therapist comfort. Each of these devices has
a duty cycle of approximately 10%, meaning 2 minutes of operation within a 20 minute
period. These devices have a thermal limiting switch to prevent overheating. As the thermal
limiting switch is engaged, the motion will slowly come to a stop. If this condition
occurs, allow the device to rest and it will automatically reset itself after a short period
of time.
THERMAL LIMITING SWITCH