II. Preparations before Use 

 

1. Take out protective foam blocks filled around the measurement sensor of the moisture meter ( “the 
meter” for short), and meanwhile confirm that the sensor is free of any substance, otherwise the meter 
will not work properly; 
2. Open the cover of battery compartment at the bottom of the meter, load in four 5# alkaline (1.5V) 
dry batteries (rechargeable batteries must not be used) according to polarity indication, or plug into AC 
power source (200V±10%, 50Hz) with AC adapter which is provided together with the meter;     
3. Put the meter on a wind-free and shake-proof horizontal platform and make the funnel closely 
attached to sample container. 
4. Get ready the sample to de tested: conduct preliminary sample screening, get rid of impurities, and 
put the sample in the meter for as long as possible to achieve temperature balance;     
5. Check the list to identify category code: for ease of operation, the calibration parameters for 
representative categories have been preset during shipment of the meter. During application, all you 
need to do is check the “category-code list” attached to this User’s Manual and chooses the 
corresponding category code for moisture measurement. 
Note: when the meter is used for trade settlement and safe storage which needs higher precision, we 
strongly recommend that you adopt standard sample of the category to be measured and conduct 
calibration and error correction of the meter so as to ensure measuring precision and safeguard your 
interest. 
 

III. Moisture Measurement 

 

1. Press down the power switch and the meter starts self-check. If no abnormality is found, the category 
code will be displayed; 

 

Note

 

 

测量

measure

 

2. Press the “ ” or “ ” key to choose the code of category to be measured (refer to the attached list 

for details); 

 

3. Put the sample into the blanking cylinder to the lower edge of funnel for later use (see Figure A); 
4. Make the blanking cylinder closely attached to the meter’s sensor, support the blanking cylinder with 
one hand and press gently the switch for discharge gate (shown in Figure B) to make the sample fully 
and evenly fall into the measuring sensor. Without pressing any key, the meter will automatically start 
the measurement and the moisture value will be displayed after the decimal point flickers several times; 

 

Note

   

样品放至下沿口

Put the sample to the lower edge

 

A

Figure A 

B

Figure B 

 

Summary of Contents for LDS-IG

Page 1: ... ISO9001 2000 certified LDS IG Computerized Moisture Meter User s Manual Quick accurate convenient AC DC power supply compatible High brightness backlight LCD clear and energy saving Multipoint calibration error correction Low power consumption automatic shutdown Automatic weighing and temperature compensation Volume weight conversion display Shanghai QingPu Oasis Inspection Technologies Co Ltd ...

Page 2: ... Lower edge of blanking cylinder the place for sample to be placed during measurement Funnel Sample container Switch for discharge gate Discharge gate Schematic drawing of LCD panel display when the dry battery is under voltage Indicate that the value is the average of the measure Indicate different working statuses of the moisture meter Indicate the corresponding unit of measurement result Indica...

Page 3: ...ry code list attached to this User s Manual and chooses the corresponding category code for moisture measurement Note when the meter is used for trade settlement and safe storage which needs higher precision we strongly recommend that you adopt standard sample of the category to be measured and conduct calibration and error correction of the meter so as to ensure measuring precision and safeguard ...

Page 4: ...lue from the standard value For instance the moisture value measured by the meter is 13 6 but actual moisture value to be displayed is 14 0 So correction value is 0 4 which means that the calibration is 0 4 upward If the correction value is negative that means the downward calibration is necessary 2 Enter into correction status take the sample out of the meter press and hold the Category key and r...

Page 5: ... code of the category to be calibrated the previous calibration parameter under this code will be overlaid 4 Enter into calibration status press and hold the 确定 OK key approx 5 6s and release it after hearing the buzz The character Calibration flickers at the bottom left corner of the LCD panel and is displayed which means that the meter has entered into the calibration status and the first standa...

Page 6: ...en press and hold the Category key until the buzz is heard The meter displays flickering which means the default parameter has been restored Shut down the meter and exit the restoration status VII Status Display The meter has self check functionality upon startup which can give prompt signals according to different working status shows that there is a sample in the sensor during startup or somethi...

Page 7: ...g put into the meter for measurement After showing the moisture value press the 确定 OK key to display the sample weight Then press the 确定 OK key again to starts automatic conversion shortly after which the volume weight of the sample will be displayed e g This value of course is for reference only 5 Communication interface used for special occasions Please contact us for its technical parameters an...

Page 8: ...racy due to unexpected factors which can be calibrated in the following manner a Horizontally place the meter in shutdown state and take everything out of the sensor Press and hold the Category key then press the Power key for startup and the meter will buzz Release the Category key and the meter will display a figure indicating the meter is under balance calibration status b Press the 确定 OK key a...