5. Close the discharge gate of the blanking cylinder, empty the sensor of the sample and prepare for the 
next measurement. 
6. In order to reduce measurement error, pay attention to the consistency in operating method: the same 
sample (especially large-grain samples such as corn) shall be subject to several measurements, whose 

results will be averaged: press the “

” key once and the average value of previous measurements 

will be shown. 

Note

    “

OK 

 

IV. Error Correction 

 

Due to objective factors in regions and category difference, preset calibration parameters with the meter 
shipment may have limitation and measurement error is likely to occur. The moisture value can be 
corrected with the following method so as to guarantee or improve measuring precision: 
1. Determine error correction value: as a rule, the moisture value measured through 105  standard 

drying oven method is adopted as the standard value. The correction value is obtained by deducting the 
measured value from the standard value. For instance, the moisture value measured by the meter is 
13.6%, but actual moisture value to be displayed is 14.0%. So correction value is +0.4, which means 
that the calibration is 0.4 upward. If the correction value is negative, that means the downward 
calibration is necessary; 
2. Enter into correction status: take the sample out of the meter, press and hold the “Category” key and 
release it after hearing the buzz. By this time, the character “Correction” flickers on the LCD panel, 
which means the meter has been in correction status, and the set error correction value (the default 
correction value before shipment is 0.0) is also displayed; 
3. Correct error: press the “ ” key to raise the correction value (e.g.,

 0.4), and then press the “

”(OK) key to save the value. The meter will flicker for confirmation. Turn off the meter or press the 

“Category” key to exit from correction status. 
The entire correction process is shown below: 

 

Note

 

 

品种

Category   

听到蜂鸣声后松开

Release it after hearing the buzz 

显示数

Indicated value 

修正

Correction   

Press the “ ” key

     

修改为

Corrected as   

修正:

Correction       

按一次确定键

Press the “

确定

”(OK) key one time     

闪烁显示

Flickering display   

修正

Correction 

 
 

Summary of Contents for LDS-IG

Page 1: ... ISO9001 2000 certified LDS IG Computerized Moisture Meter User s Manual Quick accurate convenient AC DC power supply compatible High brightness backlight LCD clear and energy saving Multipoint calibration error correction Low power consumption automatic shutdown Automatic weighing and temperature compensation Volume weight conversion display Shanghai QingPu Oasis Inspection Technologies Co Ltd ...

Page 2: ... Lower edge of blanking cylinder the place for sample to be placed during measurement Funnel Sample container Switch for discharge gate Discharge gate Schematic drawing of LCD panel display when the dry battery is under voltage Indicate that the value is the average of the measure Indicate different working statuses of the moisture meter Indicate the corresponding unit of measurement result Indica...

Page 3: ...ry code list attached to this User s Manual and chooses the corresponding category code for moisture measurement Note when the meter is used for trade settlement and safe storage which needs higher precision we strongly recommend that you adopt standard sample of the category to be measured and conduct calibration and error correction of the meter so as to ensure measuring precision and safeguard ...

Page 4: ...lue from the standard value For instance the moisture value measured by the meter is 13 6 but actual moisture value to be displayed is 14 0 So correction value is 0 4 which means that the calibration is 0 4 upward If the correction value is negative that means the downward calibration is necessary 2 Enter into correction status take the sample out of the meter press and hold the Category key and r...

Page 5: ... code of the category to be calibrated the previous calibration parameter under this code will be overlaid 4 Enter into calibration status press and hold the 确定 OK key approx 5 6s and release it after hearing the buzz The character Calibration flickers at the bottom left corner of the LCD panel and is displayed which means that the meter has entered into the calibration status and the first standa...

Page 6: ...en press and hold the Category key until the buzz is heard The meter displays flickering which means the default parameter has been restored Shut down the meter and exit the restoration status VII Status Display The meter has self check functionality upon startup which can give prompt signals according to different working status shows that there is a sample in the sensor during startup or somethi...

Page 7: ...g put into the meter for measurement After showing the moisture value press the 确定 OK key to display the sample weight Then press the 确定 OK key again to starts automatic conversion shortly after which the volume weight of the sample will be displayed e g This value of course is for reference only 5 Communication interface used for special occasions Please contact us for its technical parameters an...

Page 8: ...racy due to unexpected factors which can be calibrated in the following manner a Horizontally place the meter in shutdown state and take everything out of the sensor Press and hold the Category key then press the Power key for startup and the meter will buzz Release the Category key and the meter will display a figure indicating the meter is under balance calibration status b Press the 确定 OK key a...

Reviews: