7. Continue calibration: continue calibration in the above procedures. After the calibration of the fourth
standard sample is completed, the meter will automatically exit from the calibration status; if there is
no third or fourth standard sample, you may press the “Category” key or shut down the meter to exit
from the calibration status;
8. Recalibrate standard sample: if the measuring error is
≤
0.5%, it means the success of the calibration.
If the measuring error is bigger, recalibration is necessary.
Note: in case of any mis-operation during calibration, press and hold the “Category” key and
release it only after hearing the buzz. Return to the initial calibration status and restart the
calibration.
VI. Restoration of Calibration Data with Shipment
If you want to restore the meter’s default calibration data with shipment, the following operations shall
be performed:
Choose the code of the category to be restored. Press the “
确定
”(OK) key, and press again and hold the
“
确定
”(OK) key. Release it after hearing the buzz, and then press and hold the “Category” key until the
buzz is heard. The meter displays flickering
which means the default parameter has
been restored. Shut down the meter and exit the restoration status.
VII. Status Display
The meter has self-check functionality upon startup, which can give prompt signals according to
different working status:
: shows that there is a sample in the sensor during startup or something wrong with the
meter. The sample shall be taken out or the meter shall be inspected;
: shows that a fault occurs to the meter’s moisture measuring circuit,
temperature measuring circuit and weighing circuit respectively;
: shows that the moisture difference between standard samples is less than one percentage
point during calibration;
: shows that during calibration, there is something wrong with the sequence of standard
samples’ moisture value in terms of size;
flickering display of
at the top left corner of LCD panel: shows that the battery voltage is
insufficient and timely replacement is required;
There is a constant display of
at the top left corner of LCD panel, and meanwhile
appears on the LCD panel: shows that the batteries have been used up and the meter will be
automatically shut down in 30 seconds.