OMC-042 / 043 / 044 Data loggers
Users
’ manual
page 17
Next type 7 for Analog sensors
Choose 1 for Port 1 (mA) and the next screen will appear:
Assign a name to the sensor (option 1: Name)
1. Set the power switch to enabled and enter the warm-up time. (the power switch will supply the sensor
with 12 Volts, and will be activated the time you specify, before a measurement is taken)
2. At default Set the sample interval
to the ‘Data log interval’ so only one sample per data log interval
will be taken. You can set the sample interval to a specific value in case you want to record periodic
averages (e.g. a 10 minute wind direction, see chapter: Aggregation channels) or continues alarm
sampling.
3. Set the parameter name.
4. Set both minimum (option 6) and maximum (option 7) values of your sensor at 4mA and at 20mA. If
you don’t know those values you could determine the scaling by measuring two calibration points
(option 8). Those calibration points don’t have to be at the sensors absolute minimum and maximum,
but just two different points within the range of the sensor. (e.g. a measurement at 1m water level and
a measurement at 2m water level, while the sensors range is 0..10m)
5. If necessary option 9 can be used to perform an offset correction by measuring a single calibration
point.
6. Save and Exit
Analog sensors
[0] Exit
[1] Port 1 (mA) >> Analog1
[2] Port 2 (mA) >> Not used
[3] Port 3 (V) >> Not used
[4] Port 4 (V) >> Not used
[5] Potmeter >> Potmeter
[T] Analog input test >> Not done
>
Analog sensor
[0] Exit
[1] Name >> Analog1
[2] Sensor power switch >> Disabled
[3] Sample interval >> Data log interval
[4] Port mode >> Port 1; 4-20 mA
[5] Parameter settings >> Analog
[6] Parameter value at 4 mA >> 0 units
[7] Parameter value at 20 mA >> 100 units
[8] Determine linear conversion function (2 calibration points)
[9] Determine linear offset only (1 calibration point)
[R] Remove
>