background image

 

 

Sample Experiments 

USB2000 Operating Instructions 

26

 

 

Relative Irradiance Experiments  

Irradiance is the amount of energy at each wavelength emitted from a radiant sample. In relative terms, it is a 
comparison of the fraction of energy the sample emits and the energy the sampling system collects from a lamp 
with a blackbody energy distribution (normalized to 1 at the energy maximum). OOIBase32 calculates relative 
irradiance with the following equation… 

 

S

λ

 - D

λ

I

λ

 = B

λ

(

R

λ

 - D

λ

)

 

…where 

B

λ

 is the relative energy of the reference (calculated from the color temperature) at wavelength 

λ

S

λ

 is 

the sample intensity at wavelength 

λ

D

λ

 is the dark intensity at wavelength 

λ

R

λ

 is the reference intensity at 

wavelength 

λ

 

 

Figure 6-4:  Typical Relative Irradiance Setup. A light source with a known color temperature (such as the LS-1or  

LS-1-LL - lower right) is used to take a reference spectrum. The light to measure (lower left) is accumulated through a 

CC-3 Cosine Corrector (or FOIS integrating sphere) into an input fiber, which carries the light information to the 

spectrometer. The spectrometer then transmits the information to the PC, which compares the measured spectra against 

the reference spectrum, thus removing wavelength-dependent instrument response from the measurement. 

 

Common applications include characterizing the light output of LEDs, incandescent lamps, and other radiant 
energy sources such as sunlight. Relative irradiance measurements also include fluorescence measurements, 
which measure the energy given off by materials that have been excited by light at shorter wavelengths.  

Summary of Contents for USB2000

Page 1: ...727 733 2447 Fax 727 733 3962 8 a m 8 p m Mon Thu 8 a m 6 p m Fri EST Ocean Optics B V Europe Geograaf 24 6921 EW DUIVEN The Netherlands Phone 31 0 26 3190500 Fax 31 0 26 3190505 E mail Info OceanOpt...

Page 2: ......

Page 3: ...consent of Ocean Optics Inc in any form of binding or cover other than that in which it is published Trademarks Microsoft Windows Windows 95 Windows 98 Windows ME Windows NT Windows 2000 Windows XP a...

Page 4: ...ectrometer Specifications 7 System Compatibility 8 10 pin Accessory Connector Pinout 8 10 Pin Accessory Connector Pinout Diagram 9 10 Pin Accessory Connector Pin Definitions 9 10 Pin J2 Accessory Conn...

Page 5: ...avelength of the USB2000 31 About Wavelength Calibration 31 Calibrating the Spectrometer 31 Preparing for Calibration 31 Calibrating the Wavelength of the Spectrometer 32 Saving the New Calibration Co...

Page 6: ...sized PCs Ocean Optics OOIBase32 software application installed and configured for use with the USB2000 OOIPS2000 software if using a palm sized PC Consult the Configuring the USB2000 in OOIBase32 se...

Page 7: ...elements USB2000 Spectrometer OOIBase32 operating software Light source Sampling optics varying depending on application requirements How Sampling Works The following list explains the function of Oce...

Page 8: ...zed PC comes with 5 VDC power supply required when connecting to serial port Shipment Components The following information and documentation ships with the USB2000 Spectrometer Packing List The packin...

Page 9: ...instructions for the OOIBase32 Spectrometer Operating Software at http www oceanoptics com technical operatinginstructions asp Note You can find instructions for the OOIPS2000 Operating Software for...

Page 10: ...illustrates the movement of light through the optical bench of the USB2000 Spectrometer You can customize the items marked with an asterisk when ordering the USB2000 Note The optical bench has no mov...

Page 11: ...owards the Grating of the spectrometer Light enters the spectrometer passes through the SMA Connector Slit and Filter and then reflects off the Collimating Mirror onto the Grating 5 Grating The Gratin...

Page 12: ...ure 5 10 25 50 100 or 200 mm wide slits or fiber no slit Order sorting filters Installed longpass and bandpass filters Focal length f 4 42 mm input 68 mm output Optical resolution 0 3 10 0 nm FWHM dep...

Page 13: ...8 Me 2000 XP operating system when connecting the USB2000 to a PC via USB port or Any 32 bit version of Windows when connecting the USB2000 to a PC via serial port Compatibility for Palm sized PCs In...

Page 14: ...essory connector Pin Description 1 VUSB or 5V in 2 RS232 Tx 3 RS232 Rx 4 Lamp Enable 5 Continuous Strobe 6 Ground 7 External Trigger In 8 Single Strobe 9 I 2 C SCL 10 I 2 C SDA 10 Pin J2 Accessory Con...

Page 15: ...e USB2000 4 Insert the rectangular end of the cable into the USB port of the PC If you installed OOIBase32 prior to connecting the USB2000 the Add New Hardware Wizard appears and installs the USB2000...

Page 16: ...lt Spectrometer Configuration File screen prompts you to select a spectrometer configuration SPEC file for use with the USB2000 The unique serial number of the USB2000 precedes the file extension for...

Page 17: ...n menu Consult Chapter 4 Troubleshooting for information on identifying serial ports 4 Select the speed at which the USB2000 will operate from the Baud Rate drop down menu 115 200 baud is recommended...

Page 18: ...IBase32 stores this information for future use once you close the program Note For information on using the USB2000 with the OOIPS2000 Operating Software for the palm sized PC visit http www oceanopti...

Page 19: ...ese procedures do not correct your device driver problem you will need to obtain the Correcting Device Driver Issues document from the Ocean Optics website Visit the following web address http www oce...

Page 20: ...en Windows Explorer 2 Navigate to the Windows INF directory If the INF directory is not visible you will need to disable the Hide System Files and Folders option on in Windows Folder Options Note If t...

Page 21: ...the Hide System Files and Folders and Hide File Extensions for Known File Types options in Windows Folder Options You can access Windows Folder Options from Windows Explorer under the Tools Folder Opt...

Page 22: ...to the USB2000 s COM port 4 Verify that the COM port to which the USB2000 is interfaced is active If the ports on the PC are not labeled and you do not know the COM port number use trial and error to...

Page 23: ...on parameters using the Acquisition Parameters dialog bar or select Spectrum Configure Data Acquisition from the menu If you have followed the previous steps and started OOIBase32 the spectrometer is...

Page 24: ...scale in scope mode by Decreasing the integration time Attenuating the light going into the spectrometer Using a smaller diameter fiber Using a neutral density filter with the correct optical density...

Page 25: ...acts with the sample The output fiber carries light from the sample to the spectrometer top center which is connected to the PC far left Absorbance is also proportional to the concentration of the sub...

Page 26: ...nd merely stores a dark spectrum in memory You must select File Save Dark from the menu to permanently save the spectrum to disk Note If possible do not turn off the light source when taking a dark sp...

Page 27: ...setup The light source far right sends light via the input leg of a transmission probe into a container bottom center The light interacts with the sample The output leg of the transmission probe carri...

Page 28: ...ark from the menu bar This command merely stores a dark spectrum in memory You must select File Save Dark from the menu to permanently save the spectrum to disk Note If possible do not turn off the li...

Page 29: ...percentage R relative to the reflection from a standard reference substance S D R R D x 100 where S is the sample intensity at wavelength D is the dark intensity at wavelength R is the reference inte...

Page 30: ...by clicking the Store Dark Spectrum icon on the toolbar or by selecting Spectrum Store Dark from the menu bar This command merely stores a dark spectrum in memory You must select File Save Dark from t...

Page 31: ...avelength Figure 6 4 Typical Relative Irradiance Setup A light source with a known color temperature such as the LS 1or LS 1 LL lower right is used to take a reference spectrum The light to measure lo...

Page 32: ...ence from the menu bar to permanently save the spectrum to disk 5 Block the light path to the spectrometer Then take a dark spectrum by clicking the Store Dark Spectrum icon on the toolbar or by selec...

Page 33: ...ing the Time Acquisition Channel Configuration Screen 1 Perform the following steps on the Time Acquisition Channel Configuration screen a Select Enabled to set the time acquisition calculation for th...

Page 34: ...re Acquisition Screen section below Configuring the Time Acquisition Configuration Screen 1 Select Time Acquisition Configure Configure Acquisition from the menu bar to open the Time Acquisition Confi...

Page 35: ...the delay preceding the time acquisition process Keep in mind that the delay countdown does not begin until you start the time acquisition process Be sure to select Hours Minutes Seconds or Milliseco...

Page 36: ...coefficient nm pixel C2 is the second coefficient nm pixel2 and C3 is the third coefficient nm pixel3 You will be calculating the value for I and the three Cs Calibrating the Spectrometer Preparing fo...

Page 37: ...n of this worksheet place the observed pixel number In the third column calculate the pixel number squared and in the fourth column calculate the pixel number cubed True Wavelength nm Pixel Pixel 2 Pi...

Page 38: ...n output similar to the one shown below Regression Statistics Multiple R 0 999999831 R Square 0 999999663 Adjusted R Square 0 999999607 Standard Error 0 125540214 Observations 22 Intercept Coefficient...

Page 39: ...he Destination Location screen appears 6 Accept the default installation location or click the Browse button to specify a directory Then click the Next button The Program Manager Group screen appears...

Page 40: ...PROM memory chip on the USB2000 when using the serial mode To Save Wavelength Calibration Coefficients Using the Serial Mode 1 Open the OOIBase32 application 2 Select Spectrometer Configure from the O...

Page 41: ...al Acquisition Mode Value in Variable Delay Mode Integration time User controlled 5 ms fixed Strobe delay 0 s 2 511 s 3 s Jitter 25 ns 25 ns Using the USB2000 FLG 1 Install the OOIBase32 software on y...

Page 42: ...gnal is a rising edge trigger signal that is TTL High for the entire delay period See Figure B 2 for a timing diagram The USB2000 FLG is now configured for operation in Variable Delay Mode Figure B 1...

Page 43: ...ocusing Mirror 6 G Grating 6 I Installed Filter 6 Installing 10 L L2 Detector Collection Lens 6 Lens 6 M memory chip 1 Mirror 6 O OOIBase32 11 Overview 1 P Packing List 3 Passwords 4 Pin Definitions 9...

Reviews: