icListen User Guide
38
v2021.1
•
The instrument must now be reset in order for the update to take place. This can be done
through the web interface, or using Lucy.
Lucy: Use the “Reset” button on the “Link Setup” tab of the icListen setup panel to
reset the instrument.
Web Interface: Type the IP address of your icListen into the address bar of your
web browser (Firefox, Internet Explorer, etc), and use the “Reset” button on the
“Operations” page to reset the instrument.
Allow approximately 5 minutes for the update to complete then revisit the ABOUT page to verify
the update was successfull.
Cleaning
•
Rinse hydrophone with fresh water after every deployment.
•
This will avoid corrosion and keep salt crystals from forming on the connectors. Failure to
do this could result in increased servicing to replace connectors.
•
Clean element with warm soapy water.
•
Clean hardened biological build-up by soaking the hydrophone in a 50/50 water and 0.5
ppm acetic acid (white vinegar) solution and clean with a soft toothbrush.
•
Any accumulation of sand or mud in the connectors should be removed with spray-based
contact cleaner (isopropyl alcohol) or fresh water.
•
Failure to clean connectors can result in the splaying of the female connector and
damage to the O-ring seals.
Connector Care
•
Apply lubricant (Molykote 44 Med.) to all connectors before connecting to hydrophone.
•
If connectors are not properly lubricated connections may not seat correctly.
•
Ensure that mating surfaces of the connectors are fully seated before deployment. Failure
to do so could result in shorted connections underwater.
•
Never use excessive force to seat connectors. This may result in the pins being broken or
bent, which could result in down time while the connectors are being replaced.