background image

14

AUDIBLE ALARM

While operating in NORM or GAUG Mode the Audible will emit 1 beep per second for 

10 seconds when alarms strike, unless it is set Off. During that time, the Audible can be 

acknowledged and silenced by pressing the S button (less than 2 seconds). 

A LED Warning Light, on the lower end of the housing, is synchronized with the Audible 

and flashes as the Audible sounds. It will turn Off when the Alarm is silenced. The Audible 

and LED will not be active if the Audible is Set OFF (a group A setting).

FREE Dive Mode has its own set of Alarms which emit 3 short beeps either 1 or 3 times 

which cannot be acknowledged or set Off.

Situations that will activate the NORM/GAUG 10 second Alarm include -

** Items activate only in NORM mode.

•  Descent deeper than the Depth Alarm Set Point selected.

•  Dive Time Remaining at the Set Point selected**.

•  Elapsed Dive Time at the Set Point selected.

•  PO2 at the Set Point selected**.

•  High O2 of 300 OTU (100%)**.

•  TLBG at the Set Point selected**.

•  Ascent Rate exceeds 60 FPM (18 MPM) when deeper than 60 FT (18 M), or 30 FPM 

(9 MPM) at 60 FT (18 M) and shallower.

•  Entry into Decompression Mode (Deco)**.

•  Conditional Violation (above a required Deco Stop Depth for less than 5 minutes)**.

•  Delayed Violation (above a required Deco Stop Depth for more than 5 minutes)**.

Summary of Contents for GEO 2

Page 1: ......

Page 2: ......

Page 3: ...STEM 12 BACKLIGHT 13 AUDIBLE ALARM 14 PC INTERFACE 16 POWER SUPPLY 17 WATCH MAIN DEFAULT TIME 19 MAIN TIME 20 WATCH MENU 20 DUAL TIME 22 CDT COUNTDOWN TIMER 22 CHRONOGRAPH 25 DAILY ALARM 26 SET T MENU...

Page 4: ...50 SET A MENU NORM GAUG ALARMS 51 Set Audible Alarm 51 Set Depth Alarm 52 Set EDT Alarm 52 Set TLBG Alarm 53 Set DTR Alarm 53 Set PO2 Alarm 54 SET U MENU UTILITIES 54 Set Wet Activation 55 Set Units...

Page 5: ...O2 OTR 70 NORM DIVE MODES 71 NO DECO MAIN AND ALTS 72 DEEP STOP 73 SAFETY STOP 74 DECOMPRESSION 75 CV CONDITIONAL VIOLATION 78 DV1 DELAYED VIOLATION 1 79 DV2 DELAYED VIOLATION 2 79 DV3 DELAYED VIOLATI...

Page 6: ...Depth Alarms 99 Select M Dive Mode 100 FREE DIVE MAIN AND ALTS 101 FREE DIVE ALARMS 103 REFERENCE 105 PC INTERFACE 106 CARE AND CLEANING 108 INSPECTIONS AND SERVICE 108 BATTERY REPLACEMENT 110 ALTITUD...

Page 7: ...et depth GAUG Digital Gauge Dive Mode GLO Glow backlight HR Hour HIST History LO Low battery M Meters depth MAX Maximum MIN min Minutes time MPM Meters Per Minute NDC No Deco DTR NDL No Deco Limit NO...

Page 8: ...red trademarks trade names and service marks of Oceanic All rights are reserved PATENT NOTICE U S Patents have been issued or applied for to protect the following design features Data Sensing and Proc...

Page 9: ...9 Welcome to OCEANIC and THANK YOU for choosing the GEO 2 FEATURES FUNCTIONS AND WATCH MODE...

Page 10: ...T Descend Ascend or Stop Daily AL On Low Batt Log Mode Depth id Time or Deep Stop is triggered ICONS Surface Interval Nitrox Algorithm Time id value is Max Depth Gas value is PO2 ATA Temp value is FO2...

Page 11: ...ive Control Console utilizes 4 control buttons that allow you to maneuver through the GEO 2 s unique system of menus The buttons will be referred to as M S A and L Upper Left M mode minus Upper Right...

Page 12: ...nu Set A Menu Set M Menu Set U Menu Set FA Menu Each Menu has a start first selection and a stop last selection Upon entering a Menu movement through it begins at the start first selection then contin...

Page 13: ...ng M 2 sec while viewing the Surface Main will jump to the last menu selection such as Set T or SN then continued presses will step back through menu items BACKLIGHT To activate the Backlight at any t...

Page 14: ...wn set of Alarms which emit 3 short beeps either 1 or 3 times which cannot be acknowledged or set Off Situations that will activate the NORM GAUG 10 second Alarm include Items activate only in NORM mo...

Page 15: ...0 FT 18 M or 26 to 30 FPM 7 5 to 9 MPM at 60 FT 18 M and shallower FREE Dive Elapsed Dive Time Alarm 3 beeps every 30 seconds if set On FREE Dive Depth Alarms 1 2 3 set sequentially deeper each 3 beep...

Page 16: ...tings upload portion of the Oceanlog program can be used to set change the Main Time Date Set A group Alarms Set U group Utilities and FREE mode items using the same Interface System FO2 must be set u...

Page 17: ...t is completely disabled Battery icon shell with inner bar appears solid Fig 1a Watch and DC functions continue to be available 2 50 volts Too Low alarm level All Dive Computer operations cease and th...

Page 18: ...ntry into Surface Mode 2 50 volts Too Low alarm level Backlight is completely disabled All DC functions continue to be available during the dive Battery icon is not displayed on the dive mode screens...

Page 19: ...ed with Main Time to make it the Default Time while visiting the travel location Dual Time is the second or differential time that can be displayed simultaneously with Main Time with set On It is a Ho...

Page 20: ...M 2 sec to access the Watch Menu displaying the last selection Set T M 2 sec to access the NORM Surface Main screen S 2 sec to silence the Daily Alarm L press to activate the Backlight WATCH MENU A 2...

Page 21: ...to Watch Main DUAL TIME This selection allows Dual Time that at a travel location to be added to the Watch Main screen or removed see figure 4 ON adds shows it OFF removes it Dual Time Lead in inform...

Page 22: ...only with no toggle if no Differential has been set CDT Countdown Timer This selection provides access to a CDT for use in Watch mode Upon accessing a dive computer mode the Watch CDT opera tion will...

Page 23: ...f OFF or ON is selected revert to the Lead in screen If SET is selected access the Set CDT screen Set CDT information includes Fig 10 Graphics SEt and CDT OFF CDT hr min with Hour digits flashing A ho...

Page 24: ...then start the countdown and revert to the Lead in screen Selecting OFF will save the setting and revert to the Lead in screen The CDT will run in the background while in Watch mode until it counts do...

Page 25: ...on includes Fig 13 Graphics Chro and WATCH S 2 sec to access Chrono Status A 2 sec to step forward to Daily Alarm Lead in M 2 sec to step back to CDT Lead in Chrono Status information includes Fig 14...

Page 26: ...efault Time selected sound the Audible at the Time set every day not sound the Audible while operating in DC Modes run in the background until set Off Daily Alarm Lead in information includes Fig 15 G...

Page 27: ...icon no icon if 24 Hour Format A hold to scroll upward through Hour Set Points at a rate of 8 per second from 0 to 23 in increments of 1 hr A 2 sec to step upward through Hour Set Points one at a time...

Page 28: ...y Alarm Lead in Selecting and saving OFF will retain the time set and revert to the Lead in It can then be turned On when ready SET T MENU TIME Sequence Lead in Date Format Hour Format Default Time Du...

Page 29: ...sec to save Set Point and access Set Hour Format S 2 sec to step back to Set T Lead in Set Hour Format information includes Fig 22 Hou Format establishes number of hours displayed for Time of Day 1 to...

Page 30: ...w icon is used only while in Watch mode to indicate that Away Time is selected as the Watch Default Time in the center A 2 sec to toggle between HOME and AWAY S 2 sec to save Set Point and access Set...

Page 31: ...icon if Away is the Watch Default Time no icon if Home is the Default A hold to scroll upward through the Set Points at a rate of 8 per second from 23 through 0 to 23 in increments of 1 hour A 2 sec...

Page 32: ...our Set Points at a rate of 8 per second from 12 AM to 11 PM or 0 to 23 if 24 Hour Format in increments of 1 hr A 2 sec to step upward through Set Points one at a time M 2 sec to step back through Set...

Page 33: ...s at a rate of 8 per second from 2009 to 2052 in increments of 1 A 2 sec to step upward through Set Points one at a time M 2 sec to step back through Set Points one at a time S 2 sec to save the Year...

Page 34: ...ints at a rate of 8 per second from 1 to 31 max in increments of 1 A 2 sec to step upward through Set Points one at a time M 2 sec to step back through Set Points one at a time S 2 sec to save the Dat...

Page 35: ...on ascent to 2 FT 0 6 M for 1 second During the first 10 minutes after a NORM GAUG dive or 1 minute after a FREE dive the Dive Main screen will remain on displaying Max Depth and Elapsed Dive together...

Page 36: ...TCH DFLT TIME M 2 sec GAUG SURF MAIN M 2 sec A 2 sec Shown is access to NORM from Watch then to GAUG from NORM Access from NORM to FREE GAUG to FREE GAUG to NORM FREE to NORM and FREE to GAUG are simi...

Page 37: ...37 NORM SURFACE MENU MODES...

Page 38: ...con if Conservative Factor is set On Gas 1 icon default in Surface Mode TLBG with icon if any after a NORM or FREE dive Battery icon if voltage is low A 2 sec to access ALT 1 A hold to scroll forward...

Page 39: ...previous dive conducted A 2 sec to access ALT 2 A hold to scroll forward through Menu items M 2 sec to step back to Main L press to activate Backlight NORM SURF ALT 2 information includes Fig 31 Temp...

Page 40: ...to Fly is a countdown timer that begins counting down from 23 50 to 0 00 hr min 10 minutes after surfacing from a dive NORM GAUG or FREE Time to Dsaturate also a countdown timer provides calculated t...

Page 41: ...saturate still remains at the end of 24 hours any remaining time will be cleared FLY SAT information includes Fig 33A B Graphic FLY with Time to Fly hr min if no dive yet Graphic SAT with Time to Desa...

Page 42: ...de Refer to tables in back Plan Lead in information includes Fig 34A B Graphic PLAN PO2 alarm value set ATA with PO2 icon blank if Air Gas 1 FO2 Set Point graphic Air or numeric value 21 to 100 with F...

Page 43: ...escent and ascent rates of 60 FPM 18 MPM If less then 1 minute time is available dashes will be displayed for time and Depth values will flash PDPS information includes Fig 35A B Plan Depth value with...

Page 44: ...activated After the post dive 24 hour period has elapsed and the unit shuts off the first dive of the next activation period will be 1 In the event that a dive s elapsed time EDT exceeds 599 min the...

Page 45: ...ens from the most recent to the oldest recorded A hold to scroll through Preview screens from the most recent to the oldest recorded at a rate of 8 per second M 2 sec to step through Preview screens i...

Page 46: ...S 2 sec to access Data 2 for that dive S 2 sec to revert to Preview screen L press to activate Backlight Log Data 2 information includes Fig 39 Log Mode book icon Temperature minimum that dive with i...

Page 47: ...evert to the Data 2 screen L press to activate Backlight SET F MENU NORM FO2 Sequence FO2 Gas 1 FO2 Gas 2 50 Default When the FO2 50 Default is set Off FO2 will remain set at the last FO2 Gas 1 Set Po...

Page 48: ...ata such as PO2 O2 will not be displayed at any time during the dive on the surface or during the PDPS MODs Max Operating Depths will not be displayed on the FO2 set screen internally it will keep tra...

Page 49: ...numeric FO2 Set Point value if Nitrox flashing with FO2 icon A hold to scroll upward through Set Points at a rate of 8 per second from Air default to 21 through 100 in increments of 1 The scroll will...

Page 50: ...ond from Air default to 21 through 100 in increments of 1 The scroll will stop when A is released or at 32 50 and 80 even if A is held depressed Pressing and holding A again will resume the scroll thr...

Page 51: ...remain as set until changed Set A Lead in information includes Fig 45 Graphics SEt and A A 2 sec to step forward to Set U Lead in M 2 sec to step back to Set F Lead in S 2 sec to access Set Audible Al...

Page 52: ...2 sec to save the setting and access Set EDT Alarm S 2 sec to revert to Set Audible Alarm Set EDT Alarm information includes Fig 48 Graphics SEt and EDT Time value flashing with DIVE and min icons A...

Page 53: ...ess Set DTR Alarm S 2 sec to revert to Set EDT Alarm Set DTR Alarm information includes Fig 50 Graphics SEt and DTR Time Value min flashing with min icon A hold to scroll upward through Set Points at...

Page 54: ...rough Set Points one at a time S 2 sec to save the setting and revert to Set A Lead in S 2 sec to revert to Set DTR Alarm SET U MENU UTILITIES Sequence Lead in Wet Units DS SS Algo CF Glo SR Items app...

Page 55: ...ion includes Fig 54 Graphic Set Set Point IMP or MET flashing with FT or M icon A 2 sec to toggle between IMP and MET S 2 sec to save the setting and access Set DS S 2 sec to revert to Set Wet Set Dee...

Page 56: ...is selected operation shall access Set Algorithm If SS ON is selected Fig 56C Stop Depth and Time is displayed with the min and sec icons Time flashing A 2 sec to toggle Stop Time Set Points between...

Page 57: ...calculations for Plan and DTR values The setting locks in for 24 hours after NORM dives Set Conservative Factor CF information includes Fig 58 Graphics SEt and CF with CF icon Set Point graphics ON o...

Page 58: ...ime the backlight will remain On after L is released 0 no additional time Set Sampling Rate SR information includes Fig 60 Graphics SEt and SR Time Set Point flashing with sec icon A 2 sec to step upw...

Page 59: ...to Set U Lead in S 2 sec to access Set Dive Op Mode Set Dive Operating Mode information includes Fig 62 Graphics SEt and OP with DIVE icon Set Point flashing A 2 sec to step forward through Set Point...

Page 60: ...E icons 0 if no dive yet A 2 sec to step forward to Serial Number M 2 sec to step back to Set M Lead in S 2 sec to access History 2 History 2 information includes Fig 64 Graphic SEA or EL2 to EL7 high...

Page 61: ...r GEO 2 requires factory service Serial Number information includes Fig 65 Graphic r1A or higher indicating the revision level of the firmware GEO 2 s operating software Graphic SN with the factory pr...

Page 62: ...62...

Page 63: ...63 DIVE MODE FEATURES...

Page 64: ...e TLBG represents your relative No Deco Fig 66a or Deco status Fig 67a The first 4 segments represent No Deco status and the fifth indicates a Deco condition As your Depth and Elapsed Dive Time increa...

Page 65: ...ent Too Fast a VARI NORM GAUG The VARI Fig 68a provides a visual representation of ascent speed i e an ascent speedometer The segments represent two sets of speeds which change at a reference depth of...

Page 66: ...ADI RDP It imposes restrictions for repetitive Deco dives consid ered more risky PZ Pelagic Z performance is based on Buhlmann ZHL 16c It features NDLs that are considerably more conservative especial...

Page 67: ...3 M below or ascend 10 FT 3 M above the calculated Stop Depth for 10 seconds during the countdown the No Deco Main will replace the DS Main display and the DS feature will be disabled for the remainde...

Page 68: ...he SS Depth for 10 seconds prior to completing it the SS will be canceled for the remainder of that dive There is no Penalty if you surface prior to completing the SS or ignore it If set for Timer On...

Page 69: ...l tissue compart ments The rates each of these compartments absorb and release nitrogen is mathematically modeled and compared against a maximum allowable nitrogen level Whichever one is closest to th...

Page 70: ...than the NDC calculations for the dive will be controlled by O2 and OTR will be displayed as DTR on the Main Fig 72a identified by the O2 and min icons a Fig 71 NO DECO ALT 1 a Fig 72 NO DECO MAIN PO...

Page 71: ...71 NORM DIVE MODES...

Page 72: ...ec to acknowledge alarms L press to activate Backlight Upon ascending to 2 FT 0 6 M during a dive Surface Interval time will be displayed with the SURF icon flashing for the first 10 minutes and NDC w...

Page 73: ...co Alt 2 only if Nitrox information includes Fig 76 NX icon O2 with O2SAT icons Current PO2 value ATA with PO2 icon FO2 Set Point for Gas in use with FO2 icon Gas tank icon gas in use 1 or 2 5 sec or...

Page 74: ...ly A 2 sec to access ALTs M 2 sec to access Gas Switching S 2 sec to acknowledge alarms L press to activate Backlight DS features up to 3 ALT displays which are similar to the No Deco Main ALT1 and AL...

Page 75: ...e alarms S 2 sec to start stop Timer S 2 sec when Timer is stopped to reset it to 0 00 SS features up to 3 ALT displays which are similar to the No Deco Main ALT1 and ALT2 displays respectively Timer...

Page 76: ...top Depth indicated You should stay slightly deeper than the required Stop Depth indicated until the next shal lower Stop Depth appears Then you can slowly ascend to but not shallower than that indica...

Page 77: ...sec if A not pressed Deco Stop Alt 2 information includes Fig 84 Time of Day hr min Temperature with icon and graphic F or C A 2 sec to access ALT 2 if Nitrox Revert to Main in 5 sec if A not pressed...

Page 78: ...th both Arrows will be on solid If you descend deeper than the required Deco Stop before 5 minutes elapse Deco operation will continue with no off gassing credit given for time above the Stop Instead...

Page 79: ...ble Down Arrow icon continues to flash until descent to below required Stop Depth then full Stop icon will be on solid DV 2 DELAYED VIOLATION 2 If the calculated Deco obligation requires a Stop Depth...

Page 80: ...ontinue to be displayed as dashes for the remainder of that dive The Log for that dive will also display dashes for Max Depth VGM VIOLATION GAUGE MODE During NORM dives operation will enter VGM when D...

Page 81: ...VGM Dive Main will remain on display for 10 minutes with Surface Interval Time displayed in place of Cur rent Depth with the SURF icon flashing The graphic VIO will also still be displayed flashing Op...

Page 82: ...60 only When PO2 partial pressure of oxygen increases to the Warn ing level the audible sounds the Up Arrow icon will flash and the PO2 value will flash in place of max Depth until the audible is sil...

Page 83: ...se that apply A 2 sec to access ALTs similar to those for No Deco M 2 sec to access Gas Switching L press to activate Backlight High PO2 during Deco Fig 94 The PO2 alarm setting does not apply when in...

Page 84: ...reaches the Alarm level the audible sounds and the Up Arrow icon and the O2 value will flash in place of DTR until on the surface Fig 96 S 2 sec to acknowledge alarm A 2 sec to access ALTs similar to...

Page 85: ...e Upon ascent to 2 FT 0 6 M for 1 second surfacing the Dive Main screen is displayed for 10 minutes with access to the Dive ALTs allowed If O2 is 100 the value will flash on the Main until it is 100 t...

Page 86: ...Gas in use Preview Fig 98 M 2 sec again to access other Gas Preview similar S 2 sec to Switch to other Gas after a 3 second delay Operation reverts to the Main in 10 seconds if M or S is not pressed G...

Page 87: ...87 DIGITAL GAUGE MODE...

Page 88: ...2 sec to access SN L press to activate Backlight Upon surfacing during dives the Dive Main will remain on display for the first 10 minutes with SI in place of Depth after which the Surface Main will b...

Page 89: ...reens the Gauge Surface Menu provides access to most other selections that are similar to those described previously for NORM Mode Refer to pages 40 through 61 for descriptions of those menu items But...

Page 90: ...L press to activate Backlight Once the Run Timer is added to the Main and started it can be re moved from the Main and will continue running in the background until it is again added to the Main It ca...

Page 91: ...screen is bypassed if Max Depth is on the Main Current Depth with FT or M icon Max Depth with MAX and FT or M icons EDT Elapsed Dive Time with DIVE and min icons VARI while ascending A 2 sec to acces...

Page 92: ...d Current Depth and Max Depth will only indicate 3 dashes signifying that you are too deep Fig 107 MOD is Max Operating Depth Refer to the Specifications in the back Upon ascending above the MOD Curre...

Page 93: ...93 FREE DIVE MODE...

Page 94: ...M 2 sec to access Set M Lead in L press to activate Backlight Upon surfacing during dives the Dive Main will remain on dis play for the first 1 minute with SI in place of Depth after which the Surface...

Page 95: ...E SURF MENU Button operations A 2 sec step forward through Menu items A hold scroll forward through Menu items M 2 sec step back through Menu items S press activate the SmartGlo Backlight M 2 sec or 2...

Page 96: ...ng A 2 sec to step up through OFF ON SEt Fig 113 M 2 sec to step back through OFF ON SEt S 2 sec to save the setting If On is saved and a CDT was set the CDT will start counting down and operation wil...

Page 97: ...from 0 to 59 in increments of 1 min A 2 sec to step upward through Set Points one at a time M 2 sec to step back through Set Points one at a time S 2 sec to save the Minute Set Point and flash the Sec...

Page 98: ...to step back to CDT Lead in S 2 sec to access Set EDT Alarm EDT ALARM Factory set for a fixed 30 seconds the EDT Elapsed Dive Time alarm sounds the audible every 30 seconds while underwater in FREE D...

Page 99: ...f 8 per second from OFF to 30 to 330 FT 10 to 100 M in increments of 10 FT 1 M A 2 sec to step upward through Set Points one at a time M 2 sec to step back through Set Points one at a time S 2 sec to...

Page 100: ...y described for NORM Refer to page 59 SHARED SETTINGS To change items that FREE Mode shares with NORM Mode access the NORM Menu then Set U then Wet Activation Units Algorithm Conservative Factor Glo D...

Page 101: ...M during a dive Surface Interval time will be displayed with the SURF icon flashing for the first 1 minute and NDC will be displayed as 2 dashes Fig 120 A 2 sec to access Dive ALTs L press to activat...

Page 102: ...r ON flashing S 2 sec to toggle between ON and OFF Start or Stop the countdown and revert to Main A 2 sec to access ALT 2 Revert to Main in 10 sec if S or A is not pressed L press to activate Backligh...

Page 103: ...e the graphic CDT will flash on the Main in place of Temperature Fig 123 FREE EDT Alarm When set On the EDT alarm activates every 30 seconds during a dive The audible will sound during which time the...

Page 104: ...all 5 TLBG segments the Up Arrow icon and the graphic VIO in place of Temperature will flash and NDC will display 0 When the audible is silent the TLBG and NDC digits are re moved The graphic VIO and...

Page 105: ...105 REFERENCE...

Page 106: ...A group FREE Alarms The FO2 and Operating Mode settings must be entered using the button controls Information available for retrieval download from the GEO 2 to the PC Download portion of the program...

Page 107: ...commended is to print those sections of HELP that you consider appropriate for your interface activities PC requirements IBM or compatible Personal Computer with USB Port Intel Pentium 200 MHz or bett...

Page 108: ...and buttons are free of debris or obstructions To dissolve salt crystals use lukewarm water or a slightly acidic bath 50 white vinegar 50 fresh water After removal from the bath place the GEO 2 under...

Page 109: ...l be erased during factory service Package it using a protective cushioning material Include a legible note stating the specific reason for return your name address daytime phone number serial number...

Page 110: ...g to ensure they are not cracked or damaged If there is any sign of moisture in the GEO 2 DO NOT attempt to use it for diving until it receives proper service by an authorized Oceanic facility Battery...

Page 111: ...ring To ensure proper sealing O ring replacement is highly recommended each time the Battery is replaced Closely examine the threads of the Battery Cover and Housing for any signs of damage that migh...

Page 112: ...d from an Authorized Oceanic Dealer Use of any other O ring will void the warranty Carefully place the Battery Cover with O ring into position on the rim of the Battery Compartment then press it evenl...

Page 113: ...pressure as the current Altitude No adjustments are made during any time that the Wet Contacts are bridged When diving in high altitude waters from 3 001 to 14 000 feet 916 to 4 270 meters the GEO 2...

Page 114: ...h pressures that competitive type Free diving can expose a diver to ADDITIONAL WARNINGS Ensure that you know which Operating Mode is selected NORM GAUG or FREE prior to commencing any dive Conducting...

Page 115: ...115 TECHNICAL DATA...

Page 116: ...0 04 0 04 0 03 0 03 0 03 0 03 0 03 0 03 0 03 0 03 190 0 04 0 04 0 04 0 03 0 03 0 03 0 03 0 03 0 03 0 03 0 03 0 00 PZ ALGORITHM NDLS HR MIN AT ALTITUDE METRIC Altitude 0 916 1221 1526 1831 2136 2441 27...

Page 117: ...0 05 0 04 0 04 0 04 0 04 0 04 0 04 0 03 0 03 0 03 190 0 05 0 04 0 04 0 04 0 04 0 04 0 04 0 03 0 03 0 03 0 03 0 03 DSAT ALGORITHM NDLS HR MIN AT ALTITUDE METRIC Altitude 0 916 1221 1526 1831 2136 2441...

Page 118: ...Altitude corrections and O2 limits based on NOAA tables OPERATIONAL PERFORMANCE Function Accuracy Depth 1 of full scale Timers 1 second per day Dive Mode Activation Must be in Dive Computer mode if We...

Page 119: ...ation and every 1 000 feet 305 meters higher Power 1 3 vdc CR2430 Lithium battery Panasonic or equivalent Shelf life Up to 7 years when shipped from factory in deep sleep mode Replacement User annual...

Page 120: ...second Dual Time of Day 00 00 to 23 59 hr min 1 minute Dual Time Differential 23 to 0 to 23 59 hr 1 hour Watch Countdown Timer 23 59 to 0 00 hr min 1 minute Watch Chrono Lap Time 0 00 to 1 59 59_99 0...

Page 121: ...Stop Time 5 00 to 0 00 min sec 1 second Safety Stop Run Timer 0 00 to 9 59 min sec 1 second GAUG Dive Run Timer 0 to 999 min 1 minute Deco Stop Time 0 to 999 min 1 minute Total Ascent Time 0 to 999 m...

Page 122: ...______ Firmware Rev ____________________________________________ Date of Purchase ____________________________________________ Purchased from ____________________________________________ Below to be f...

Page 123: ...Nice France Tel 33 0 4 93 72 43 00 Fax 33 0 4 93 72 43 05 E mail info subaquadis fr OCEANIC ITALY Genova Italy Tel 39 010 545 1212 Fax 39 010 518 4232 Web www oceanicitalia com E mail info oceanicital...

Page 124: ......

Reviews: