3-11. IO/Local
To configure the power supply for the RS-232, RS485 and TCP/IP interface.
RS232C, RS485 interface is standard, in case of TCP/IP, it is optional.
When the power supply is shipped, I/O interface is set the RS232C protocol and baud rate is 9600bps.
Communication setting is able to be set by front panel key.
▌
The I/O and local setting are stored in non-volatile memory, and does not change
when power has been off or after a remote interface reset.
▌
If remote interface works, the lamp of RMT on front panel lits and power supply is remotely
controlled preferentially.
▌
If you want to control power supply on local mode, at first finish remote interface
and then press the "I/O & LOCAL" key. Lamp of "RMT" turn off and you can contorol.
Although the RMT lamp is off but you can not control the in strument, maybe it is under KEY LOCk.
Press the "Protection" key and unlock.
RS232C Setting
RS232C setting procedure.
▌
Turn on the power supply
After turned on, check the power supply displays"**OUTPUT OFF**" Mode.
▌
Press "I/O" key in order to set RS232C interface.
I/O> 1.RS232C
Or it differs following before state
▌
Following before state, appears RS232C, RS485 and TCP/IP on the
display. If the display is not "1.RS232C" press the cursor key until changing
to "I/O> 1.RS232C" on the display
I/O> 1.RS232C
▌
Press "I/O" key in order to set baud rate.
LCD Display
>br 1: 9600 bps Or it may differ depending on previous state.
▌
Baud rate is consist of 9600, 19200, 38400, 57600, 115200 and 230400bps.
You must match between remote interface and device's baud rate.
For examples, in other to select 115200bps, press right cursor key four times
LCD Display
>br 5: 115200 bps
▌
In order to finish the I/O setting, press the "I/O" key
LCD Display
SAVED
Accordingly, Remote interface is set as RS232C, Baud Rate is 115200bps
bps ? - The initials of Bit per Second, data transmission unit of 1bit per one second.
- 40 -
Note
Summary of Contents for EX 100-12
Page 1: ...Programmable DC Power Supply User Manual www odacore com EX Series Manual Part NO 019EX 03...
Page 4: ...Safety Symbol...
Page 91: ...92...