EX Series Operating Manual
6-5. AUTO CURSOR MV
▌If the power supply doesn't receive any operation by using front-panel within a certain period
of time, the power supply's cursor location go down to minimum changing unit automatically
when you set the "AUTO CURSOR MV" mode is enable.
▌After passed long time, if a operator turns encoderder knob by mistake, cursor location of
voltage and current is at minimum unit position, so it can reduce the DUT damage.
▌F
ti
▌Function.
> DISABLE
Disable "AUTO CURSOR MV" mode.
> ENABLE
Enable "AUTO CURSOR MV" mode.
AUTO CURSOR MV(Moving) Operating
▌Power switch on in a pressing "FACTORY" key.
▌Power switch ON
▌When appearing "1. user-mem clear" message, release "FACTORY" key.
In a pressing
▌Enter to "3. AUTO CURSOR MV" mode by using cursor key.
▌In other to set, press the "FACTORY" key one more time.
LCD Display
disable
Release
▌Set the "enable" mode by using cursor key.
LCD Display
enable
▌To finish the "AUTO CURSOR" mode, press the "FACTORY" key
one more time.
LCD Display
done
LCD Display
done
≫ Related remote interface command
FACT:AUTO-CUR {DIS|ENA}
FACT:AUTO-CUR?
Return Value(0,1)
A
l
FACT AUTO CUR?
R
l
1
"AUTO CURSOR"
d i
bl
Apply:
FACT:AUTO-CUR?
Return value : 1
"AUTO CURSOR" mode is enable state
FACT:AUTO-CUR DIS
"AUTO CURSOR" mode changed to
disable state