1-6. Check After Power On
When turning on the power switch, the front-panel display will light up briefly while the
instrument performs its power-on default value setting. And also keep the ex-remote
interface setting mode, voltage value is zero and current value is max value.
OVP & OCP will be set on maximum output value.
Display Procedure on LCD
▌
Homepage
"
WWW.ODACORE.COM
" will be displayed.
Visit our homepage and we will provide various service and info about upgrading, manual and software.
▌Also, Front lamp OVP, OCP, LOCK, RMT will be all lighted.
CV & CC are not included.
▌"
INITIALIZING. . .
" Message will be displayed.
There will be a reset using non-volatile memory.
▌Also, Front lamp OVP, OCP, LOCK, RMT lights will be turned off.
CV & CC are not included.
▌"
SELF-TESTING. . .
" Message will be displayed.
While self-testing, message below will be displayed.
Ch
k F
t
l'
ti
F
t
l T
t
Check Front panel's connection
Remote interface Test
Check Remote interface for PC network
Memory data verification Test
Check Instrument's info & setting
ADC H/W error Test
Check ADC controlling
UnRegulated condition Test
Check Output voltage Floating
ADC/DAC Calibration Test
Check ADC & DAC Calibration Data
When an error occurs, there will be a alarm sound from instrument, and error No will be stored.
Front panel Test
Press Key to check error code.
Please read “7. Error Messages” to check details about error code.
Default Setting Value
▌OVP : OVP Max value, OVP setting ON ▌Remote Interface : maintain previous setting
▌OCP : OCP Max value, OCP setting ON ▌
Condition after Self-testing
: "
**OUTPUT OFF**
"
▌Output Voltage : 0V
▌Cursor Location : Default Voltage Select
▌Output Voltage : 0V
▌Cursor Location : Default Voltage Select
▌Output Current : Limit setting max value
Voltage : 1V scale
▌KEY LOCK : OFF
Current : 5A > 100mA scale
50A > 1A scale
The RS232C is attached in the instrument when the power supply is shipped from the
Note1
p
pp y
pp
factory for remote interface configuration and baudrate is set 9600bps at first time.
In case of choosing RS485 interface, address no. is 05 when this is shipped from the factory.
It has the option function that is last setup stated memory & recall. This function, when you
turn down the instrument, the device store the last state(voltage, current and state) and after
then the operator turn on again this starts from last state
Note2
then the operator turn on again, this starts from last state.