4. CALIBRATION
Do not use calibrate function without person with knowledge.
Warning
Calibration should be done
> every 180 days for accuracy
cyclically.
> every 365 days to be activated properly
4-1. Attributes
▌Calibrate without opening case
▌Calibration using Front panel Key
▌PC based calibration from Remote Interface
▌PC based calibration from Remote Interface
▌Store in fixed memory
▌Support Calibration data Backup & Value Backup
4-2. Preparing for accurate calibration
▌Use measuring instruments should measure spec of power supply.
▌Give more than 1 hour in temp 20~30 degree to warm up.
▌Humidity should be below 80%
▌Humidity should be below 80%.
▌Connect Power Supply's output terminal to load instrument.
▌Other instruments that causes magnetic field should be off.