Operating Current CALIBRATION
▌Setting current mode after sending electric load switch ON command.
▌Send power supply current minimum calibration command.
CAL CURR MIN
Command :
▌Set maximum current value to make CC mode.
▌After specific time measure DVM Voltage connected with current monitoring resist
CAL:CURR MIN
▌After specific time, measure DVM Voltage connected with current monitoring resist.
▌Send output current value from calculating program to power supply value.
For instance, if calculated value is 0.1234, send as below :
Command
CAL:CURR 0.1234
▌Send Voltage Maximum Calibration command to Power Supply.
Command
▌After specific time, measure DVM Voltage connected with current monitoring resist.
▌Send output value from calculating program to power supply value.
CAL:CURR MAX
For instance, if calculated value is 10.123, send as below :
Command
CAL:CURR 10.123
- 58 -