5. FACTORY
This instrument supports user's memory backup, Cycling Mode, Calibration backup & restore op
5-1. Atrribute
▌Reset 10 data in User's Memory.
▌Reset data from CYCLING MODE.
▌Reset data from CYCLING MODE.
▌Restore & Backup CALIBRATION Data.
▌Saves into fixed memory.
▌Controlling using Remote Interface.
5-2. FACTORY KEY STRUCTURE
RECALL /
FACTORY
POWER
ON
1. CYCLING
5-3. CYCLING CLEAR
FACTORY
ON
2. USER-MEM
3. CALI -
4. CALI -
5 3. CYCLING CLEAR
▌Command that resets memory 1~100 from Cycling mode.
▌When the command above is done, can not be returned.
▌Reset Description
>Step Voltage 00.000V
>Step Current 00.000A
>Step Slope T 00000.0s
>Step Delay T 00000.0s
>Step Delay T 00000.0s
>Sequence St Set No.1 memory
>Sequence Fi Set memory for No.100
>Repeat
100 Times
Operating CYCLING CLEAR
▌Power Switch ON while holding FACTORY Key.
▌
g
y
Holding
▌Power Switch ON
▌
When
"
1. CYCLING CLEAR
" Message is displayed, stop holding Key.
▌To clear Cycling Memory press FACTORY Key once more
▌To clear Cycling Memory, press FACTORY Key once more.
▌
When
"
MEMORY RESET
" is displayed, It is properly activated.
▌"
SYSTEM RESTART
" will be displayed and reset the system.
≫ Related Remote Interface Command
FACTory:CYCling CLEar
Refer: FACT:CYC CLE Cycling Memory Reset
- 59 -