5-6. CALI-BACKUP
▌When forwarding, fixed data will be saved in protected non-volatile memory.
▌To assure accuracy, instrument should be calibrated within 6 months.
To be protected safely, There could be a calibration data backup.
▌When the backup is operated, previous backup description cannot be restored.
Operating CALI-BACKUP
▌Switch Power ON while holding FACTORY Key.
Holding
▌Power Switch ON
▌
When
"
1. CYCLING CLEAR
" is displayed, stop holding FACTORY Key.
▌P
i ht
k
th
ti
t
t
CALI B
k
▌Press right cursor key three times to enter CALI-Backup.
VFD Display Description
▌To backup calibration data, press FACTORY Key once more.
4. CALI - BACKUP
▌
When
"
BACKUPED
" is displayed, it is activated properly.
▌"
SYSTEM RESTART
" will be displayed and reset the system.
≫ Related Remote Interface Command
Factory:Calibration Backup
Refer: FACT:CAL BAC Calibration Back-up