-30, "En route to cal the curr"
This occurs when you have sent current calibration commands while voltage calibration.
-31, "Over curr min parameter"
When Current Min value is out of range.
Read reference "4-5. CALIBRATE setting range"
g
g
-32, "Under curr max parameter"
This error occurs when Max current value range is out of lowest value.
Read reference "4-5. CALIBRATE Setting Range"
-33, "Over curr max parameter"
This error occurs when Current Max value range is out of highest value.
g
g
Read reference "4-5. CALIBRATE Setting Range"
-34, "Not allowed command under cal"
Other command cannot be used while Remote Calibration.
7-4. Cycling Mode Error
7 4. Cycling Mode Error
-50, "Invalid sequence of cycling mode"
Occurs when Sequence's finish number is bigger than start number.
ex) Start Number Input 10
Finish NumberInput 05, and press RUN/STOP Key to sense error code.
7-5. Calibration Error
Readback Calibration will be activated while calibrating.
Check whether there is an error occurred.
-74, "ADC-V low limit over"
Occurs when voltage is out of ADC Low range.
-75, "ADC-V high limit over"
Occurs when voltage is out of ADC High range.
-76, "ADC-A low limit over"
Occurs when current is out of ADC Low range.
77 "ADC A hi h li it
"
-77, "ADC-A high limit over"
Occurs when current is out of ADC High range.
- 77 -