Compact / Series 7600 Installation Procedure
15
System Operational Verification
Generator
mA
kV
S
min s
HIP
1.
Press the C-Up and C-Down buttons and verify that the C-arm lifts
smoothly from the top to the bottom limits.
2.
Press the Auto Fluoro button and verify that the LED illuminates. Press
the x-ray button and verify that the kV automatically servos by monitoring
the kV display and that an image appears on the monitor. Verify that the
mA display displays the mA value while x-rays are being taken.
3.
Press the Pulse Fluoro button and verify that the LED illuminates. Press
the x-ray button and verify that the system produces pulsed x-rays and that
an image appears on the monitor.
4.
Press the Snapshot button and verify that the LED illuminates. Press the
x-ray button and verify that an exposure initiates and then automatically
terminates. The image should remain on the monitor until you take the next
exposure.